Identifying and Quantifying Relative Concentrations of Epimers in Mixtures via Cyclic Ion Mobility Mass Spectrometry: Dexamethasone and Betamethasone as a Case Study

Epimers can show different biological activities and different pharmacological behaviors; therefore, their separation and analysis are crucial in the drug development process. Due to their similar chemical and physical properties, separation of epimers is challenging. This study demonstrates the application of cyclic ion mobility-mass spectrometry to separate, identify, and quantify dexamethasone and betamethasone in a binary mixture. Cyclic IMS separation of the isolated protonated dimer resulted in three peaks: dexamethasone homodimer, betamethasone homodimer, and their heterodimer. Besides providing improved separation over the protonated monomer, the presence of a heterodimer peak provides additional confirmation of an isomeric mixture. We identified the dexamethasone and betamethasone homodimer peaks by infusing pure solutions of each epimer and measuring each pure homodimer's arrival time. The measured peak areas indicated that the heterodimer is formed at twice the rate of each homodimer and that dexamethasone and betamethasone contribute equally to the heterodimer signal. Using this observation, we could accurately calculate the relative concentrations of each epimer by adding half of the heterodimer peak area to each homodimer peak area. These findings enable the identification and quantification of dexamethasone and betamethasone based on the arrival time distributions of their protonated dimers. This is the first demonstration of accurate relative quantification of epimers by separating charged dimers in the gas phase.

Authors

Institutions

Publication Details

Journal
Journal of the American Society for Mass Spectrometry
Published
2024-08-26
DOI
https://doi.org/10.1021/jasms.4c00258
Citations
6
Primary Topic
Mass Spectrometry Techniques and Applications
Type
article
Field-Weighted Citation Impact
1.44

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Identifying and Quantifying Relative Concentrations of Epimers in Mixtures via Cyclic Ion Mobility Mass Spectrometry: Dexamethasone and Betamethasone as a Case Study

Kenneth W. Lee, David V. Dearden, Sudam S. Mane
6 citations
Journal of the American Society for Mass Spectrometry
Mass Spectrometry Techniques and Applications
1.44
article

Identifying and Quantifying Relative Concentrations of Epimers in Mixtures via Cyclic Ion Mobility Mass Spectrometry: Dexamethasone and Betamethasone as a Case Study

Kenneth W. Lee, David V. Dearden, Sudam S. Mane
article en
6 citations

Abstract

Epimers can show different biological activities and different pharmacological behaviors; therefore, their separation and analysis are crucial in the drug development process. Due to their similar chemical and physical properties, separation of epimers is challenging. This study demonstrates the application of cyclic ion mobility-mass spectrometry to separate, identify, and quantify dexamethasone and betamethasone in a binary mixture. Cyclic IMS separation of the isolated protonated dimer resulted in three peaks: dexamethasone homodimer, betamethasone homodimer, and their heterodimer. Besides providing improved separation over the protonated monomer, the presence of a heterodimer peak provides additional confirmation of an isomeric mixture. We identified the dexamethasone and betamethasone homodimer peaks by infusing pure solutions of each epimer and measuring each pure homodimer's arrival time. The measured peak areas indicated that the heterodimer is formed at twice the rate of each homodimer and that dexamethasone and betamethasone contribute equally to the heterodimer signal. Using this observation, we could accurately calculate the relative concentrations of each epimer by adding half of the heterodimer peak area to each homodimer peak area. These findings enable the identification and quantification of dexamethasone and betamethasone based on the arrival time distributions of their protonated dimers. This is the first demonstration of accurate relative quantification of epimers by separating charged dimers in the gas phase.

Journal of the American Society for Mass SpectrometryVol. 35(10)
Brigham Young University (US)
Department of Chemistry and Biochemistry, Brigham Young University
Openalex Percentile: Top 20%
Mass Spectrometry Techniques and Applications
1.44
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.