Interocular Symmetry and Intermachine Reproducibility of Optic Disc and Macular Parameters Measured by Two Different Models of Optical Coherence Tomography

Purpose: To compare the interocular symmetry and investigate the intermachine reproducibility of optic disc and macular data measured by spectral-domain high-definition optical coherence tomography (HD-OCT) Cirrus HD-OCT 4000 and HD-OCT 5000 from healthy subjects. Patients and Methods: Forty-three volunteers were examined with both HD-OCT 4000 and HD-OCT 5000 at the same visit. Optic nerve head (ONH) and macular data were acquired using ONH Cube 200× 200 scans and macular volume cube 512× 128 scans, respectively. Results: The average age of the participants was 33 ± 8.6 years. Interocular OCT parameters of ONH and macula showed a high correlation between the right and left eyes regardless of HD-OCT models, displaying a low coefficient of variation (CV). However, the average retinal nerve fiber layer (RNFL) was thicker (96.67± 11.19μm vs 95.3± 10.89μm, p< 0.01), and the average central subfield thickness (261.51± 17.45μm vs 262.51± 17.39 μm, p< 0.01) and cube average thickness (283.91± 13.59μm vs 286.55± 13.09μm, p< 0.05) were thinner when measured by Cirrus 4000 compared to 5000. Intermachine reproducibility and reliability of RNFL and macular parameters exhibited a high intraclass correlation coefficient (ICC) (0.985) and low CV (2.4%). Ganglion cell-inner plexiform layer (GCIPL) measured by two OCT models showed similar values with an average thickness of 85 μm and had high intermachine reproducibility with high ICC (0.993) and low CV (1.2%). Conclusion: High interocular symmetry was observed across both HD-OCT models. Intermachine reproducibility for RNFL and all macular parameters was also high. GCIPL showed minimal intermachine differences with high reproducibility and reliability. Thus, the results imply that GCIPL values measured by two Cirrus OCT models may be used interchangeably. Keywords: neuro imaging, optical coherence tomography, interocular symmetry, intermachine reproducibility, retina, ganglion cell-inner plexiform layer, interchangeability A Letter to the Editor has been published for this article . A Response to Letter by Dr Galvan has been published for this article.

Authors

Institutions

Publication Details

Journal
Clinical ophthalmology
Published
2024-08-01
DOI
https://doi.org/10.2147/opth.s465360
Citations
2
Primary Topic
Glaucoma and retinal disorders
Type
article
Field-Weighted Citation Impact
1.08

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Interocular Symmetry and Intermachine Reproducibility of Optic Disc and Macular Parameters Measured by Two Different Models of Optical Coherence Tomography

Yu-Min Huang, Lívia Figueiredo Pereira, Salma Firdaus, Ge Yang
2 citations
Clinical ophthalmology
Glaucoma and retinal disorders
1.08
article

Interocular Symmetry and Intermachine Reproducibility of Optic Disc and Macular Parameters Measured by Two Different Models of Optical Coherence Tomography

Yu-Min Huang, Lívia Figueiredo Pereira, Salma Firdaus, Ge Yang
article en
2 citations

Abstract

Purpose: To compare the interocular symmetry and investigate the intermachine reproducibility of optic disc and macular data measured by spectral-domain high-definition optical coherence tomography (HD-OCT) Cirrus HD-OCT 4000 and HD-OCT 5000 from healthy subjects. Patients and Methods: Forty-three volunteers were examined with both HD-OCT 4000 and HD-OCT 5000 at the same visit. Optic nerve head (ONH) and macular data were acquired using ONH Cube 200× 200 scans and macular volume cube 512× 128 scans, respectively. Results: The average age of the participants was 33 ± 8.6 years. Interocular OCT parameters of ONH and macula showed a high correlation between the right and left eyes regardless of HD-OCT models, displaying a low coefficient of variation (CV). However, the average retinal nerve fiber layer (RNFL) was thicker (96.67± 11.19μm vs 95.3± 10.89μm, p< 0.01), and the average central subfield thickness (261.51± 17.45μm vs 262.51± 17.39 μm, p< 0.01) and cube average thickness (283.91± 13.59μm vs 286.55± 13.09μm, p< 0.05) were thinner when measured by Cirrus 4000 compared to 5000. Intermachine reproducibility and reliability of RNFL and macular parameters exhibited a high intraclass correlation coefficient (ICC) (0.985) and low CV (2.4%). Ganglion cell-inner plexiform layer (GCIPL) measured by two OCT models showed similar values with an average thickness of 85 μm and had high intermachine reproducibility with high ICC (0.993) and low CV (1.2%). Conclusion: High interocular symmetry was observed across both HD-OCT models. Intermachine reproducibility for RNFL and all macular parameters was also high. GCIPL showed minimal intermachine differences with high reproducibility and reliability. Thus, the results imply that GCIPL values measured by two Cirrus OCT models may be used interchangeably. Keywords: neuro imaging, optical coherence tomography, interocular symmetry, intermachine reproducibility, retina, ganglion cell-inner plexiform layer, interchangeability A Letter to the Editor has been published for this article . A Response to Letter by Dr Galvan has been published for this article.

Clinical ophthalmologyVol. Volume 18
Linköping University (SE), Pontifícia Universidade Católica de Minas Gerais (BR), Universitas Airlangga (ID), Huizhou Central People's Hospital (CN), He Eye Hospital (CN)
Linköpings Universitet, Faculty of Medicine and Health, University of Sydney
Zero hunger
Openalex Percentile: Top 23%
Glaucoma and retinal disorders
1.08
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.