Case Report: Extraction of a stylet-driven lead for left bundle branch area pacing >2 years after implantation

Left bundle branch pacing has recently emerged as a significant alternative to right ventricular pacing. The rate of implanted stylet-driven septal leads is expected to increase substantially in the coming years, along with the need to manage long-term complications. Experience in extracting these leads is currently very limited; however, the number of complex extractions is anticipated to increase in the future. We report a complex case involving the extraction of a long-dwelling Solia lead used for left bundle branch pacing in a 21-year-old man. The lead was extracted through the implant vein 27 months after implantation, using a methodology that involved a locking stylet and compression coil. The new lead insertion was challenging due to venous occlusion but after successful venoplasty, the His lead was successfully implanted. The postoperative course was uneventful, demonstrating the feasibility of extraction without complications.

Authors

Institutions

Publication Details

Journal
Frontiers in Cardiovascular Medicine
Published
2024-08-26
DOI
https://doi.org/10.3389/fcvm.2024.1457025
Citations
2
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.80
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Case Report: Extraction of a stylet-driven lead for left bundle branch area pacing >2 years after implantation

David Židan, Ivana Grgić Romić, Ana Lanča Bastiančić, Sandro Brusich et al.
2 citations
Frontiers in Cardiovascular Medicine
Cardiac pacing and defibrillation studies
0.80
article

Case Report: Extraction of a stylet-driven lead for left bundle branch area pacing >2 years after implantation

David Židan, Ivana Grgić Romić, Ana Lanča Bastiančić, Sandro Brusich, Mate Mavrić
article en
2 citations

Abstract

Left bundle branch pacing has recently emerged as a significant alternative to right ventricular pacing. The rate of implanted stylet-driven septal leads is expected to increase substantially in the coming years, along with the need to manage long-term complications. Experience in extracting these leads is currently very limited; however, the number of complex extractions is anticipated to increase in the future. We report a complex case involving the extraction of a long-dwelling Solia lead used for left bundle branch pacing in a 21-year-old man. The lead was extracted through the implant vein 27 months after implantation, using a methodology that involved a locking stylet and compression coil. The new lead insertion was challenging due to venous occlusion but after successful venoplasty, the His lead was successfully implanted. The postoperative course was uneventful, demonstrating the feasibility of extraction without complications.

Frontiers in Cardiovascular MedicineVol. 11
University of Rijeka (HR)
Openalex Percentile: Top 24%
Cardiac pacing and defibrillation studies
0.80
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines