Late-onset left atrial appendage occlusion device–related thrombus attributed to mitral bioprosthetic stenosis: a case report
This case demonstrates that increased stasis and left atrial dysfunction from prosthetic mitral stenosis can be a risk factor for late DRT after successful LAAO. The use of a LAAO occlusion device in the presence of a mitral bioprosthesis requires more frequent echocardiographic monitoring to assess both the function of the prosthesis and a delayed formation of thrombus. More studies need to be conducted to assess the safety of percutaneous LAAO devices in those with mitral bioprosthesis.