Biodesign: Engineering an aortic endograft explantation tool

Endovascular aortic repair (EVAR) graft failure can be as high as 16% to 30% owing to endoleak, graft migration, or infection, often necessitating explantation, leading to potential morbidity (31%) and mortality (6.3%). Graft prongs frequently tear through the endothelium during explantation, leading to endothelial damage and subsequent fatal bleeding. The current standard of care involves different suboptimal techniques such as the syringe technique in which a cylinder is improvised by cutting a syringe in half and pushed over the graft hooks in a rotating motion, until covered for manual explantation. Because there is no commercially available product to address this shortcoming in graft explantation, we engage in the biodesign process to produce a functional explantation device. We designed and prototyped multiple potential solutions to remove EVAR endografts safely. Silicone tubing with EVAR endografts deployed in the lumen were used to simulate a grafted aorta and test each prototype. Prototypes were compared in their ability to meet design criteria including decrease in graft diameter, prevention of arterial dissection, ease of use, and decrease in procedure time. After determining the single best prototype, surgeon feedback was elicited to iteratively improve the original design. The most effective design uses a tapered lumenal geometry that decreases the EVAR graft diameter and uses stainless steel beads to prevent shear stress to the simulated aorta. A distal grip allows for easy single hand manipulation of the device, while a latching mechanism allows for smooth placement and removal over the endograft. After rigorous prototyping, our device proved feasible and effective for safe EVAR explantation, allowing this procedure to be performed safely.

Authors

Institutions

Publication Details

Journal
Journal of Vascular Surgery Cases and Innovative Techniques
Published
2024-08-23
DOI
https://doi.org/10.1016/j.jvscit.2024.101599
Citations
1
Primary Topic
Aortic aneurysm repair treatments
Type
article
Field-Weighted Citation Impact
0.42

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Biodesign: Engineering an aortic endograft explantation tool

Paul Haddad, Daanish Sheikh, Maham Rahimi, Solyman Hatami et al.
1 citations
Journal of Vascular Surgery Cases and Innovative Techniques
Aortic aneurysm repair treatments
0.42
article

Biodesign: Engineering an aortic endograft explantation tool

Paul Haddad, Daanish Sheikh, Maham Rahimi, Solyman Hatami, Vamsi Maturi, Alwin Mathew, S. Lu
article en
1 citations

Abstract

Endovascular aortic repair (EVAR) graft failure can be as high as 16% to 30% owing to endoleak, graft migration, or infection, often necessitating explantation, leading to potential morbidity (31%) and mortality (6.3%). Graft prongs frequently tear through the endothelium during explantation, leading to endothelial damage and subsequent fatal bleeding. The current standard of care involves different suboptimal techniques such as the syringe technique in which a cylinder is improvised by cutting a syringe in half and pushed over the graft hooks in a rotating motion, until covered for manual explantation. Because there is no commercially available product to address this shortcoming in graft explantation, we engage in the biodesign process to produce a functional explantation device. We designed and prototyped multiple potential solutions to remove EVAR endografts safely. Silicone tubing with EVAR endografts deployed in the lumen were used to simulate a grafted aorta and test each prototype. Prototypes were compared in their ability to meet design criteria including decrease in graft diameter, prevention of arterial dissection, ease of use, and decrease in procedure time. After determining the single best prototype, surgeon feedback was elicited to iteratively improve the original design. The most effective design uses a tapered lumenal geometry that decreases the EVAR graft diameter and uses stainless steel beads to prevent shear stress to the simulated aorta. A distal grip allows for easy single hand manipulation of the device, while a latching mechanism allows for smooth placement and removal over the endograft. After rigorous prototyping, our device proved feasible and effective for safe EVAR explantation, allowing this procedure to be performed safely.

Journal of Vascular Surgery Cases and Innovative TechniquesVol. 10(6)
Houston Methodist (US), The University of Texas at San Antonio Health Science Center (US)
National Science Foundation, Houston Methodist Hospital
Openalex Percentile: Top 34%
Aortic aneurysm repair treatments
0.42
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.