Simulation study of hemodynamic commonality of umbrella-shaped inferior vena cava filter using computational fluid dynamics

The inferior vena cava filter (IVCF) has been used in clinical practice for decades to prevent fatal pulmonary embolism. One of the most commonly used types of IVCFs features a conical umbrella-like design. The geometric similarity among these umbrella-shaped IVCFs suggests a potential commonality in their hemodynamic behavior. However, many previous studies have primarily analyzed the blood flow around a single umbrella-shaped IVCF, failing to capture the general hemodynamic characteristics of this filter type. This paper uses computational fluid dynamics to simulate the blood flow around five typical umbrella-shaped IVCFs, aiming to investigate their hemodynamic commonality. For all five IVCFs, prothrombotic stagnation zones are observed downstream of the filter head, extending nearly 12 times the diameter of the inferior vena cava (IVC). The “viscous block” effect of the filter wires accelerates blood flow within the conical region of the filter, enhancing the potential for clot capture and dissolution. The deployment of umbrella-shaped IVCF remarkably increases the wall shear stress (WSS) on the IVC wall, and the WSS on the filter wire continuously increases in the flow direction. Additionally, the viscous resistance of each umbrella-shaped IVCF contributes several times more to the total flow resistance than the pressure resistance. Our results suggest that alterations in the morphology of the filter wires of umbrella-shaped IVCFs significantly impact the flow resistance of the filter and the WSS distribution pattern of the IVC. These hemodynamic changes may lead to life-threatening conditions, such as IVC lesions, filter rupture, and displacement.

Authors

Institutions

Publication Details

Journal
Physics of Fluids
Published
2024-08-01
DOI
https://doi.org/10.1063/5.0217997
Citations
4
Primary Topic
Venous Thromboembolism Diagnosis and Management
Type
article
Field-Weighted Citation Impact
2.20
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Simulation study of hemodynamic commonality of umbrella-shaped inferior vena cava filter using computational fluid dynamics

Chun‐Hian Lee, Jingying Wang, Mingrui Li, Shiyue Zhang et al.
4 citations
Physics of Fluids
Venous Thromboembolism Diagnosis and Management
2.20
article

Simulation study of hemodynamic commonality of umbrella-shaped inferior vena cava filter using computational fluid dynamics

Chun‐Hian Lee, Jingying Wang, Mingrui Li, Shiyue Zhang, Yue Zhou, Song Xue
article en
4 citations

Abstract

The inferior vena cava filter (IVCF) has been used in clinical practice for decades to prevent fatal pulmonary embolism. One of the most commonly used types of IVCFs features a conical umbrella-like design. The geometric similarity among these umbrella-shaped IVCFs suggests a potential commonality in their hemodynamic behavior. However, many previous studies have primarily analyzed the blood flow around a single umbrella-shaped IVCF, failing to capture the general hemodynamic characteristics of this filter type. This paper uses computational fluid dynamics to simulate the blood flow around five typical umbrella-shaped IVCFs, aiming to investigate their hemodynamic commonality. For all five IVCFs, prothrombotic stagnation zones are observed downstream of the filter head, extending nearly 12 times the diameter of the inferior vena cava (IVC). The “viscous block” effect of the filter wires accelerates blood flow within the conical region of the filter, enhancing the potential for clot capture and dissolution. The deployment of umbrella-shaped IVCF remarkably increases the wall shear stress (WSS) on the IVC wall, and the WSS on the filter wire continuously increases in the flow direction. Additionally, the viscous resistance of each umbrella-shaped IVCF contributes several times more to the total flow resistance than the pressure resistance. Our results suggest that alterations in the morphology of the filter wires of umbrella-shaped IVCFs significantly impact the flow resistance of the filter and the WSS distribution pattern of the IVC. These hemodynamic changes may lead to life-threatening conditions, such as IVC lesions, filter rupture, and displacement.

Physics of FluidsVol. 36(8)
Shandong University (CN), Jinan Central Hospital (CN), Beihang University (CN)
Openalex Percentile: Top 13%
Venous Thromboembolism Diagnosis and Management
2.20
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.