Transhepatic Transcatheter Pulmonary Valve Replacement
Transcatheter pulmonary valve replacement (TPVR) is complicated in patients without adequate femoral or internal jugular vascular access. Transhepatic vascular access has been shown to be safe and effective across a spectrum of diagnostic and interventional procedures. Closure of the hepatic venous tract can be accomplished with a multitude of readily available vascular occlusion devices. The rates of major adverse events are low: 5% to 8% with hemoperitoneum and complete heart block are most significant. To our knowledge, this is the first report of using transhepatic access for TPVR; closure of the hepatic venous tract was achieved with an Amplatzer vascular plug type II.