Quantitative Analysis of Chondroitin Sulfate and Dermatan Sulfate Using Capillary Electrophoresis With Gaussian Distribution‐Based Peak Fitting for Improved Peak Resolution

ABSTRACT In this work, capillary electrophoresis with Gaussian distribution‐based peak fitting for improved peak resolution for simultaneous quantitative analysis of chondroitin sulfate (CS) and dermatan sulfate (DS) was developed for the first time. The effects of different pH, voltage, and peak shapes on the peak fitting were investigated. The results indicate that separation conditions that affect peak shape have a significant impact on peak fitting. Under optimized separation conditions, the peak fitting achieves high accuracy and precision, the relative standard deviation ( n = 5) of peak area and migration time of CS and DS were 1.1%, 1.8% and 0.94%, 0.84%, respectively. Furthermore, the results obtained by peak fitting were close to real values. The developed method was used to analyze the purity of CS. The results had a high consistency with the labeled value; furthermore, the peak fitting could point out the number of possible impurities. The developed method has good potential for the analysis of other glycosaminoglycans with peak overlap.

Authors

Institutions

Publication Details

Journal
Separation Science Plus
Published
2024-08-20
DOI
https://doi.org/10.1002/sscp.202300244
Citations
5
Primary Topic
Proteoglycans and glycosaminoglycans research
Type
article
Field-Weighted Citation Impact
1.63

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quantitative Analysis of Chondroitin Sulfate and Dermatan Sulfate Using Capillary Electrophoresis With Gaussian Distribution‐Based Peak Fitting for Improved Peak Resolution

Guanglian Zhou, Xiumei Liu, Kang Wang, Qianru Xu
5 citations
Separation Science Plus
Proteoglycans and glycosaminoglycans research
1.63
article

Quantitative Analysis of Chondroitin Sulfate and Dermatan Sulfate Using Capillary Electrophoresis With Gaussian Distribution‐Based Peak Fitting for Improved Peak Resolution

Guanglian Zhou, Xiumei Liu, Kang Wang, Qianru Xu
article en
5 citations

Abstract

ABSTRACT In this work, capillary electrophoresis with Gaussian distribution‐based peak fitting for improved peak resolution for simultaneous quantitative analysis of chondroitin sulfate (CS) and dermatan sulfate (DS) was developed for the first time. The effects of different pH, voltage, and peak shapes on the peak fitting were investigated. The results indicate that separation conditions that affect peak shape have a significant impact on peak fitting. Under optimized separation conditions, the peak fitting achieves high accuracy and precision, the relative standard deviation ( n = 5) of peak area and migration time of CS and DS were 1.1%, 1.8% and 0.94%, 0.84%, respectively. Furthermore, the results obtained by peak fitting were close to real values. The developed method was used to analyze the purity of CS. The results had a high consistency with the labeled value; furthermore, the peak fitting could point out the number of possible impurities. The developed method has good potential for the analysis of other glycosaminoglycans with peak overlap.

Separation Science PlusVol. 7(10)
Qilu University of Technology (CN), Shandong University (CN), Chang'an University (CN), Shandong Academy of Sciences (CN)
National Key Research and Development Program of China
Openalex Percentile: Top 18%
Proteoglycans and glycosaminoglycans research
1.63
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Quantitative Analysis of Chondroitin Sulfate and Dermatan Sulfate Using Capillary Electrophoresis With Gaussian Distribution‐Based Peak Fitting for Improved Peak Resolution — Guanglian Zhou, Xiumei Liu, et al. · Separation Science Plus (2024) | TGRS Research Map | TGRS