Feasibility of 3D MRI fingerprinting for rapid knee cartilage T<sub>1</sub>, T<sub>2,</sub> and T<sub>1ρ</sub> mapping at 0.55T: Comparison with 3T

Low-field strength scanners present an opportunity for more inclusive imaging exams and bring several challenges including lower signal-to-noise ratio (SNR) and longer scan times. Magnetic resonance fingerprinting (MRF) is a rapid quantitative multiparametric method that can enable multiple quantitative maps simultaneously. To demonstrate the feasibility of an MRF sequence for knee cartilage evaluation in a 0.55T system we performed repeatability and accuracy experiments with agar-gel phantoms. Additionally, five healthy volunteers (age 32 ± 4 years old, 2 females) were scanned at 3T and 0.55T. The MRI acquisition protocols include a stack-of-stars T<sub>1ρ</sub>-enabled MRF sequence, a VIBE sequence with variable flip angles (VFA) for T<sub>1</sub> mapping, and fat-suppressed turbo flash (TFL) sequences for T<sub>2</sub> and T<sub>1ρ</sub> mappings. Double-Echo steady-state (DESS) sequence was also used for cartilage segmentation. Acquisitions were performed at two different field strengths, 0.55T and 3T, with the same sequences but protocols were slightly different to accommodate differences in signal-to-noise ratio and relaxation times. Cartilage segmentation was done using five compartments. T<sub>1</sub>, T<sub>2</sub>, and T<sub>1ρ</sub> values were measured in the knee cartilage using both MRF and conventional relaxometry sequences. The MRF sequence demonstrated excellent repeatability in a test-retest experiment with model agar-gel phantoms, as demonstrated with correlation and Bland-Altman plots. Underestimation of T<sub>1</sub> values was observed on both field strengths, with the average global difference between reference values and the MRF being 151 ms at 0.55T and 337 ms at 3T. At 0.55T, MRF measurements presented significant biases but strong correlations with the reference measurements. Although a larger error was present in T<sub>1</sub> measurements, MRF measurements trended similarly to the conventional measurements for human subjects and model agar-gel phantoms.

Authors

Institutions

Publication Details

Journal
NMR in Biomedicine
Published
2024-08-21
DOI
https://doi.org/10.1002/nbm.5250
Citations
3
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
1.21

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Feasibility of 3D MRI fingerprinting for rapid knee cartilage T1, T2, and T1ρ mapping at 0.55T: Comparison with 3T

Anmol Monga, Ravinder R. Regatte, Hector Lise de Moura, Mahesh Keerthivasan et al.
3 citations
NMR in Biomedicine
Advanced MRI Techniques and Applications
1.21
article

Feasibility of 3D MRI fingerprinting for rapid knee cartilage T1, T2, and T1ρ mapping at 0.55T: Comparison with 3T

Anmol Monga, Ravinder R. Regatte, Hector Lise de Moura, Mahesh Keerthivasan, Xiaoxia Zhang, Marcelo V. W. Zibetti
article en
3 citations

Abstract

Low-field strength scanners present an opportunity for more inclusive imaging exams and bring several challenges including lower signal-to-noise ratio (SNR) and longer scan times. Magnetic resonance fingerprinting (MRF) is a rapid quantitative multiparametric method that can enable multiple quantitative maps simultaneously. To demonstrate the feasibility of an MRF sequence for knee cartilage evaluation in a 0.55T system we performed repeatability and accuracy experiments with agar-gel phantoms. Additionally, five healthy volunteers (age 32 ± 4 years old, 2 females) were scanned at 3T and 0.55T. The MRI acquisition protocols include a stack-of-stars T<sub>1ρ</sub>-enabled MRF sequence, a VIBE sequence with variable flip angles (VFA) for T<sub>1</sub> mapping, and fat-suppressed turbo flash (TFL) sequences for T<sub>2</sub> and T<sub>1ρ</sub> mappings. Double-Echo steady-state (DESS) sequence was also used for cartilage segmentation. Acquisitions were performed at two different field strengths, 0.55T and 3T, with the same sequences but protocols were slightly different to accommodate differences in signal-to-noise ratio and relaxation times. Cartilage segmentation was done using five compartments. T<sub>1</sub>, T<sub>2</sub>, and T<sub>1ρ</sub> values were measured in the knee cartilage using both MRF and conventional relaxometry sequences. The MRF sequence demonstrated excellent repeatability in a test-retest experiment with model agar-gel phantoms, as demonstrated with correlation and Bland-Altman plots. Underestimation of T<sub>1</sub> values was observed on both field strengths, with the average global difference between reference values and the MRF being 151 ms at 0.55T and 337 ms at 3T. At 0.55T, MRF measurements presented significant biases but strong correlations with the reference measurements. Although a larger error was present in T<sub>1</sub> measurements, MRF measurements trended similarly to the conventional measurements for human subjects and model agar-gel phantoms.

NMR in BiomedicineVol. 37(12)
U.S. National Science Foundation (US), New York University (US)
National Institutes of Health
Good health and well-being
Openalex Percentile: Top 20%
Advanced MRI Techniques and Applications
1.21
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.