Modes of administration of nitric oxide devices and ventilators flow-by impact the delivery of pre-determined concentrations
We identified three different generations of iNO-devices, based on their gas administration modalities, that were associated with highly variable iNO concentrations' accuracy. Ventilator's flow by significantly impacted iNO concentration. Only the Proportional devices permitted to accurately deliver iNO whatever the conditions and the ventilators tested.
Authors
- Arnaud Lesimple
- Yoann Zerbib (ORCID: https://orcid.org/0000-0002-9937-6096)
- Alice Vuillermoz
- Nathan Prouvez
- C. Brault (ORCID: https://orcid.org/0000-0001-6210-2270)
- François Beloncle (ORCID: https://orcid.org/0000-0002-0590-459X)
- Jean‐Christophe Richard (ORCID: https://orcid.org/0000-0003-1503-3035)
- Mathieu Le Tutour
- Jean-Marie Forel
- Alain Mercat (ORCID: https://orcid.org/0000-0003-0582-700X)
- Mary Schmitt
- M. Lefranc
Institutions
- Centre Hospitalier Universitaire Amiens-Picardie (FR)
- Hôpital Nord (FR)
- Air Liquide (France) (FR)
- Université d'Angers (FR)
Publication Details
- Journal
- Annals of Intensive Care
- Published
- 2024-08-21
- DOI
- https://doi.org/10.1186/s13613-024-01351-w
- Citations
- 4
- Primary Topic
- Respiratory Support and Mechanisms
- Type
- article
- Field-Weighted Citation Impact
- 1.68