Optimized Pediatric X-ray Doses through Automatic Tube Voltage Control in Conventional Machines

Chest radiography remains a cornerstone of pediatric pneumonia diagnosis, offering a non-invasive approach. However, minimizing radiation exposure in children is crucial. This study aimed to optimize pediatric X-ray doses by developing a prototype for an automatic tube voltage control system in conventional X-ray machines. We identified key factors influencing radiation dose during pediatric chest X-rays. Utilizing the Commission of European Communities' recommended tube voltage (>60 kVp), we analyzed the Entrance Surface Dose (ESD). Subsequently, an automatic control mechanism for tube voltage (<60 kVp) was constructed using a ZMPT101B voltage sensor, a control unit, and a patient-type selector. The ZMPT101B sensor detects preset tube voltage. A dedicated algorithm analyzes these values, and if they exceed recommended levels, the control unit adjusts them. Results are displayed on a Liquid Crystal Display (LCD). Validation of ESD with varying tube voltage and current settings confirmed that exposure time and tube voltage significantly impact ESD more than tube current. Lower kVp settings with high tube currents are unsuitable for children due to the limited penetration depth of such photons, resulting in higher tissue absorption without improving image quality. This novel monitoring system addresses this concern by automatically adjusting tube voltage for optimal image acquisition at reduced dose levels.

Authors

Publication Details

Journal
International Journal For Multidisciplinary Research
Published
2024-08-19
DOI
https://doi.org/10.36948/ijfmr.2024.v06i04.26292
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimized Pediatric X-ray Doses through Automatic Tube Voltage Control in Conventional Machines

Hakkins Raj -, Tewodros Degefe -, Tewodros Belay Alemneh -
International Journal For Multidisciplinary Research
Radiation Detection and Scintillator Technologies
article

Optimized Pediatric X-ray Doses through Automatic Tube Voltage Control in Conventional Machines

Hakkins Raj -, Tewodros Degefe -, Tewodros Belay Alemneh -
article en

Abstract

Chest radiography remains a cornerstone of pediatric pneumonia diagnosis, offering a non-invasive approach. However, minimizing radiation exposure in children is crucial. This study aimed to optimize pediatric X-ray doses by developing a prototype for an automatic tube voltage control system in conventional X-ray machines. We identified key factors influencing radiation dose during pediatric chest X-rays. Utilizing the Commission of European Communities' recommended tube voltage (>60 kVp), we analyzed the Entrance Surface Dose (ESD). Subsequently, an automatic control mechanism for tube voltage (<60 kVp) was constructed using a ZMPT101B voltage sensor, a control unit, and a patient-type selector. The ZMPT101B sensor detects preset tube voltage. A dedicated algorithm analyzes these values, and if they exceed recommended levels, the control unit adjusts them. Results are displayed on a Liquid Crystal Display (LCD). Validation of ESD with varying tube voltage and current settings confirmed that exposure time and tube voltage significantly impact ESD more than tube current. Lower kVp settings with high tube currents are unsuitable for children due to the limited penetration depth of such photons, resulting in higher tissue absorption without improving image quality. This novel monitoring system addresses this concern by automatically adjusting tube voltage for optimal image acquisition at reduced dose levels.

International Journal For Multidisciplinary ResearchVol. 6(4)
Openalex Percentile: Top 86%
Radiation Detection and Scintillator Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Optimized Pediatric X-ray Doses through Automatic Tube Voltage Control in Conventional Machines — Hakkins Raj -, Tewodros Degefe -, et al. · International Journal For Multidisciplinary Research (2024) | TGRS Research Map | TGRS