Progress toward absolute calibrations of DC x-ray sources from 1 keV to >30 keV for Z diagnostics applications

Numerous x-ray diagnostics at the Z facility, time-resolved and time-integrated, would benefit from a local calibrated x-ray source. For many years, Z has used low-power Manson x-ray sources to provide some characterization information for components of Z diagnostics below 30 keV. More recently, we have acquired TruFocus x-ray sources, which operate at voltages up to 100 kV depending on the anode material and model. These DC sources have demonstrated good stability and repeatability. Absolute calibrations of either sources or detectors usually involve another calibrated detector from another source, which has not been readily available locally. Therefore, we have adopted the strategy of characterizing a sensitive semiconductor diode detector by controlling or inferring the active volume. This characterized detector is then used to scale the Manson x-ray spectral shape as measured by an Amptek energy-resolving detector. The result agrees well with the modeled source spectrum. We can apply this technique to the TruFocus as well.

Authors

Institutions

Publication Details

Journal
Review of Scientific Instruments
Published
2024-08-01
DOI
https://doi.org/10.1063/5.0213745
Primary Topic
Electron and X-Ray Spectroscopy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Progress toward absolute calibrations of DC x-ray sources from 1 keV to >30 keV for Z diagnostics applications

Timothy Webb, Patrick W. Lake
Review of Scientific Instruments
Electron and X-Ray Spectroscopy Techniques
article

Progress toward absolute calibrations of DC x-ray sources from 1 keV to >30 keV for Z diagnostics applications

Timothy Webb, Patrick W. Lake
article en

Abstract

Numerous x-ray diagnostics at the Z facility, time-resolved and time-integrated, would benefit from a local calibrated x-ray source. For many years, Z has used low-power Manson x-ray sources to provide some characterization information for components of Z diagnostics below 30 keV. More recently, we have acquired TruFocus x-ray sources, which operate at voltages up to 100 kV depending on the anode material and model. These DC sources have demonstrated good stability and repeatability. Absolute calibrations of either sources or detectors usually involve another calibrated detector from another source, which has not been readily available locally. Therefore, we have adopted the strategy of characterizing a sensitive semiconductor diode detector by controlling or inferring the active volume. This characterized detector is then used to scale the Manson x-ray spectral shape as measured by an Amptek energy-resolving detector. The result agrees well with the modeled source spectrum. We can apply this technique to the TruFocus as well.

Review of Scientific InstrumentsVol. 95(8)
Sandia National Laboratories (US)
National Nuclear Security Administration
Affordable and clean energy
Openalex Percentile: Top 88%
Electron and X-Ray Spectroscopy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Progress toward absolute calibrations of DC x-ray sources from 1 keV to >30 keV for Z diagnostics applications — Timothy Webb, Patrick W. Lake · Review of Scientific Instruments (2024) | TGRS Research Map | TGRS