Cone-beam CT with a noncircular (sine-on-sphere) orbit: imaging performance of a clinical system for image-guided interventions
Image quality differences between circular and noncircular CBCT orbits were quantitatively evaluated on a clinical system in the context of neurosurgery. The primary performance advantage for the noncircular orbit was the improved sampling and elimination of cone-beam artifacts.
Authors
- Jeffrey H. Siewerdsen (ORCID: https://orcid.org/0000-0001-6644-1317)
- Stephen Chen (ORCID: https://orcid.org/0000-0002-5580-831X)
- Moiz Ahmad
- A. Kyle Jones
- Shaan M. Raza
Institutions
- The University of Texas MD Anderson Cancer Center (US)
Publication Details
- Journal
- Journal of Medical Imaging
- Published
- 2024-08-22
- DOI
- https://doi.org/10.1117/1.jmi.11.4.043503
- Primary Topic
- Medical Imaging Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00