Picosecond pulsed beams of light and heavy keV ions at the Time-of-Flight Medium energy ion scattering system at Uppsala University

A 16 MHz electrostatic beam chopper is implemented at the Uppsala University ToF-MEIS system to complement the 4 MHz sinusoidal scanning, refining the resolution without drift-tube bunching. The system is benchmarked with H2+, He+, and Ne+ ions in transmission and backscattering geometries. The estimated true pulse duration for 60 keV He+ is 34 ps while the direct beam impinging on the detector resulted in measured pulse widths of 295 ps for He ions and 481 ps for Ne ions. In backscattering geometries, ions impinging on the target yield measured pulse durations of 459 ns for H ions, 550 ps for He ions and 810 ps for Ne ions and lead to energy resolution measurements of 2.4 keV (100 keV H), 0.9 keV (60 keV He), and 2.4 keV (160 keV Ne). Discussions cover straggling effects on achievable energy resolution and how to obtain estimates of the true duration of the ion pulse.

Authors

Institutions

Publication Details

Journal
Nuclear Instruments and Methods in Physics Research Section B Beam Interactions with Materials and Atoms
Published
2024-08-22
DOI
https://doi.org/10.1016/j.nimb.2024.165494
Citations
5
Primary Topic
Ion-surface interactions and analysis
Type
article
Field-Weighted Citation Impact
1.87

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Picosecond pulsed beams of light and heavy keV ions at the Time-of-Flight Medium energy ion scattering system at Uppsala University

D. Wessman, Daniel Primetzhofer, Eleni Ntemou, Radek Holeňák
5 citations
Nuclear Instruments and Methods in Physics Research Section B Beam Interactions with Materials and Atoms
Ion-surface interactions and analysis
1.87
article

Picosecond pulsed beams of light and heavy keV ions at the Time-of-Flight Medium energy ion scattering system at Uppsala University

D. Wessman, Daniel Primetzhofer, Eleni Ntemou, Radek Holeňák
article en
5 citations

Abstract

A 16 MHz electrostatic beam chopper is implemented at the Uppsala University ToF-MEIS system to complement the 4 MHz sinusoidal scanning, refining the resolution without drift-tube bunching. The system is benchmarked with H2+, He+, and Ne+ ions in transmission and backscattering geometries. The estimated true pulse duration for 60 keV He+ is 34 ps while the direct beam impinging on the detector resulted in measured pulse widths of 295 ps for He ions and 481 ps for Ne ions. In backscattering geometries, ions impinging on the target yield measured pulse durations of 459 ns for H ions, 550 ps for He ions and 810 ps for Ne ions and lead to energy resolution measurements of 2.4 keV (100 keV H), 0.9 keV (60 keV He), and 2.4 keV (160 keV Ne). Discussions cover straggling effects on achievable energy resolution and how to obtain estimates of the true duration of the ion pulse.

Nuclear Instruments and Methods in Physics Research Section B Beam Interactions with Materials and AtomsVol. 556
Uppsala University (SE)
Vetenskapsrådet
Affordable and clean energy
Openalex Percentile: Top 16%
Ion-surface interactions and analysis
1.87
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Picosecond pulsed beams of light and heavy keV ions at the Time-of-Flight Medium energy ion scattering system at Uppsala University — D. Wessman, Daniel Primetzhofer, et al. · Nuclear Instruments and Methods in Physics Research Section B Beam Interactions with Materials and Atoms (2024) | TGRS Research Map | TGRS