A Modified Benchtop Test Method to Measure the Movements of Intervertebral Body Fusion Devices

ABSTRACT This study developed and evaluated a new benchtop test method to measure the movements of different designs of intervertebral body fusion devices (IBFDs) under cyclic loads. The experimental method simulates and evaluates the movement resistance of intervertebral cages under flexion-extension cyclic loads experienced by the lumbar spine. The present method modifies a method developed previously and offers a possibility to be more clinically relevant and robust. This is because it allowed for closer achievement of levels of IBFD rotation reported in the literature and introduced the ability to control the flexion-extension movement and, consequently, the rotation of the cage more precisely. Polyurethane foam blocks were used as the bone substitute material. Optical and contactless displacement measurements were performed using a robust six degrees of freedom measuring system (OptiTrack) to evaluate the movements of the IBFD during testing. The results obtained by applying the new benchtop method showed greater sensitivity in capturing the implant movements in all directions measured, with magnitude significantly superior from Ribeiro et al. (2022). A test method to measure the movements of different IBFD designs was modified to better simulate and control the lumbar spine flexion-extension movements. The proposed benchtop methodology can evaluate the IBFD movements in all directions to more efficiently discriminate in a potentially standardized test the efficacy of different IBFD designs.

Authors

Institutions

Publication Details

Journal
Journal of Testing and Evaluation
Published
2024-08-19
DOI
https://doi.org/10.1520/jte20230697
Primary Topic
Spine and Intervertebral Disc Pathology
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Modified Benchtop Test Method to Measure the Movements of Intervertebral Body Fusion Devices

Arthur Paiva Grimaldi Santos, Edison da Rosa, Hani Haider, Carlos Rodrigo de Mello Roesler et al.
Journal of Testing and Evaluation
Spine and Intervertebral Disc Pathology
article

A Modified Benchtop Test Method to Measure the Movements of Intervertebral Body Fusion Devices

Arthur Paiva Grimaldi Santos, Edison da Rosa, Hani Haider, Carlos Rodrigo de Mello Roesler, Kevin Mauricio Menon Ribeiro, Izabela Ramos de Sousa
article en

Abstract

ABSTRACT This study developed and evaluated a new benchtop test method to measure the movements of different designs of intervertebral body fusion devices (IBFDs) under cyclic loads. The experimental method simulates and evaluates the movement resistance of intervertebral cages under flexion-extension cyclic loads experienced by the lumbar spine. The present method modifies a method developed previously and offers a possibility to be more clinically relevant and robust. This is because it allowed for closer achievement of levels of IBFD rotation reported in the literature and introduced the ability to control the flexion-extension movement and, consequently, the rotation of the cage more precisely. Polyurethane foam blocks were used as the bone substitute material. Optical and contactless displacement measurements were performed using a robust six degrees of freedom measuring system (OptiTrack) to evaluate the movements of the IBFD during testing. The results obtained by applying the new benchtop method showed greater sensitivity in capturing the implant movements in all directions measured, with magnitude significantly superior from Ribeiro et al. (2022). A test method to measure the movements of different IBFD designs was modified to better simulate and control the lumbar spine flexion-extension movements. The proposed benchtop methodology can evaluate the IBFD movements in all directions to more efficiently discriminate in a potentially standardized test the efficacy of different IBFD designs.

Journal of Testing and EvaluationVol. 52(5)
Universidade Federal de Santa Catarina (BR), University of Nebraska Medical Center (US)
Openalex Percentile: Top 78%
Spine and Intervertebral Disc Pathology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A Modified Benchtop Test Method to Measure the Movements of Intervertebral Body Fusion Devices — Arthur Paiva Grimaldi Santos, Edison da Rosa, et al. · Journal of Testing and Evaluation (2024) | TGRS Research Map | TGRS