A rate-responsive duty-cycling protocol for leadless pacemaker synchronization

Dual-chamber leadless pacemakers (LLPMs) consist of two implants, one in the right atrium and one in the right ventricle. Inter-device communication, required for atrioventricular (AV) synchrony, however, reduces the projected longevity of commercial dual-chamber LLPMs by 35-45%. This work analyzes the power-saving potential and the resulting impact on AV-synchrony for a novel LLPM synchronization protocol. Relevant parameters of the proposed window scheduling algorithm were optimized with system-level simulations investigating the resulting trade-off between transceiver current consumption and AV-synchrony. The parameter set included the algorithm's setpoint for the target number of windows per cardiac cycle and the number of averaging cycles used in the window update calculation. The sensing inputs for the LLPM model were derived from human electrocardiogram recordings in the MIT-BIH Arrhythmia Database. Transceiver current consumption was estimated by combining the simulation results on the required communication resources with electrical measurements of a receiver microchip developed for LLPM synchronization in previous work. The performance ratio given by AV-synchrony divided by current consumption was maximized for a target of one window per cardiac cycle and three averaging cycles. Median transceiver current of both LLPMs combined was 166 nA (interquartile range: 152-183 nA) and median AV-synchrony was 92.5%. This corresponded to median reduction of 18.3% and 3.2% in current consumption and AV-synchrony, respectively, compared to a non-rate-responsive implementation of the same protocol, which prioritized maximum AV-synchrony. In conclusion, adopting a rate-responsive communication protocol may significantly increase device longevity of dual-chamber LLPMs without compromising AV-synchrony, potentially reducing the frequency of device replacements.

Authors

Institutions

Publication Details

Journal
Biomedical Engineering Letters
Published
2024-08-19
DOI
https://doi.org/10.1007/s13534-024-00413-z
Primary Topic
Neuroscience and Neural Engineering
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A rate-responsive duty-cycling protocol for leadless pacemaker synchronization

Adrian Ryser, Thomas Niederhäuser, Jürgen Burger, Andreas Haeberlin et al.
Biomedical Engineering Letters
Neuroscience and Neural Engineering
article

A rate-responsive duty-cycling protocol for leadless pacemaker synchronization

Adrian Ryser, Thomas Niederhäuser, Jürgen Burger, Andreas Haeberlin, Tobias Reichlin
article en

Abstract

Dual-chamber leadless pacemakers (LLPMs) consist of two implants, one in the right atrium and one in the right ventricle. Inter-device communication, required for atrioventricular (AV) synchrony, however, reduces the projected longevity of commercial dual-chamber LLPMs by 35-45%. This work analyzes the power-saving potential and the resulting impact on AV-synchrony for a novel LLPM synchronization protocol. Relevant parameters of the proposed window scheduling algorithm were optimized with system-level simulations investigating the resulting trade-off between transceiver current consumption and AV-synchrony. The parameter set included the algorithm's setpoint for the target number of windows per cardiac cycle and the number of averaging cycles used in the window update calculation. The sensing inputs for the LLPM model were derived from human electrocardiogram recordings in the MIT-BIH Arrhythmia Database. Transceiver current consumption was estimated by combining the simulation results on the required communication resources with electrical measurements of a receiver microchip developed for LLPM synchronization in previous work. The performance ratio given by AV-synchrony divided by current consumption was maximized for a target of one window per cardiac cycle and three averaging cycles. Median transceiver current of both LLPMs combined was 166 nA (interquartile range: 152-183 nA) and median AV-synchrony was 92.5%. This corresponded to median reduction of 18.3% and 3.2% in current consumption and AV-synchrony, respectively, compared to a non-rate-responsive implementation of the same protocol, which prioritized maximum AV-synchrony. In conclusion, adopting a rate-responsive communication protocol may significantly increase device longevity of dual-chamber LLPMs without compromising AV-synchrony, potentially reducing the frequency of device replacements.

Biomedical Engineering LettersVol. 14(6)
University of Bern (CH), Bern University of Applied Sciences (CH), University Hospital of Bern (CH)
Hasler Stiftung, Schweizerische Herzstiftung, Novartis Stiftung für Medizinisch-Biologische Forschung
Openalex Percentile: Top 87%
Neuroscience and Neural Engineering
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.