THE EFFECT OF CEMENTED FEMORAL COMPONENT DESIGN AND PROPHYLACTIC WIRE PLACEMENT AGAINST EARLY PERIPROSTHETIC FEMORAL FRACTURE IN PRIMARY TOTAL HIP ARTHROPLASTY: A BIOMECHANICAL STUDY

Previous studies have highlighted differences in the risk of periprosthetic fracture between tapered slip (TS) and composite beam (CB) stems. This biomechanical study explored periprosthetic fractures around these stems and the effect of adding a 16-gauge calcar or diaphyseal wire to TS stems on their resistance to torque. A power analysis determined a sample size of 7 specimens per group, assuming a standard deviation of 14.8 Nm in peak torque, to provide 90% power to detect a difference of at least 30 Nm between groups. Twenty-one TS stems (eight alone, six with calcar wiring, seven with diaphyseal wiring placed 2 cm distal to the lesser trochanter) and seven CB stems were cemented into standard Sawbones. A servo-hydraulic test machine applied a 1000 N load with a 1-degree rotation per second until failure. The peak torque at failure was measured, and the fracture location recorded. Comparisons were performed using two-sample t-tests. CB stems exhibited a significantly higher peak torque at failure (205.3 Nm) than TS stems (159.5 Nm, p=0.020). Calcar-wire-TS (148.2 Nm, p=0.036) and diaphyseal-wire-TS (164.9 Nm, p=0.036) were both weaker than CB stems. Wired-TS stems showed no significant difference from non-wired-TS stems. Additionally, the study could not conclude that calcar wiring is stronger than diaphyseal wiring. All TS fractures occurred at the mid-stem, simulating a B-type fracture, while the addition of the diaphyseal wire shifted the fracture location more distally in four of seven stems (p=0.0699). This biomechanical study supports the clinical evidence that CB stems have stronger resistance to torque than TS stems and may explain lower risk of periprosthetic fracture. The addition of calcar or diaphyseal wires to TS stems resulted in no significant changes in peak torque to fracture. In patients at high risk of periprosthetic fracture, CB cemented stems should be considered.

Authors

Institutions

Publication Details

Journal
Orthopaedic Proceedings
Published
2024-08-19
DOI
https://doi.org/10.1302/1358-992x.2024.16.045
Citations
1
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.57
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

THE EFFECT OF CEMENTED FEMORAL COMPONENT DESIGN AND PROPHYLACTIC WIRE PLACEMENT AGAINST EARLY PERIPROSTHETIC FEMORAL FRACTURE IN PRIMARY TOTAL HIP ARTHROPLASTY: A BIOMECHANICAL STUDY

Rafael J. Sierra, Ichiro Tsukamoto, Joshua R. Labott, Chunfeng Zhao et al.
1 citations
Orthopaedic Proceedings
Orthopaedic implants and arthroplasty
0.57
article

THE EFFECT OF CEMENTED FEMORAL COMPONENT DESIGN AND PROPHYLACTIC WIRE PLACEMENT AGAINST EARLY PERIPROSTHETIC FEMORAL FRACTURE IN PRIMARY TOTAL HIP ARTHROPLASTY: A BIOMECHANICAL STUDY

Rafael J. Sierra, Ichiro Tsukamoto, Joshua R. Labott, Chunfeng Zhao, Alexander W. Hooke, Sergio F. Guarin Perez, Guangyue Zhao, David J. Restrepo
article en
1 citations

Abstract

Previous studies have highlighted differences in the risk of periprosthetic fracture between tapered slip (TS) and composite beam (CB) stems. This biomechanical study explored periprosthetic fractures around these stems and the effect of adding a 16-gauge calcar or diaphyseal wire to TS stems on their resistance to torque. A power analysis determined a sample size of 7 specimens per group, assuming a standard deviation of 14.8 Nm in peak torque, to provide 90% power to detect a difference of at least 30 Nm between groups. Twenty-one TS stems (eight alone, six with calcar wiring, seven with diaphyseal wiring placed 2 cm distal to the lesser trochanter) and seven CB stems were cemented into standard Sawbones. A servo-hydraulic test machine applied a 1000 N load with a 1-degree rotation per second until failure. The peak torque at failure was measured, and the fracture location recorded. Comparisons were performed using two-sample t-tests. CB stems exhibited a significantly higher peak torque at failure (205.3 Nm) than TS stems (159.5 Nm, p=0.020). Calcar-wire-TS (148.2 Nm, p=0.036) and diaphyseal-wire-TS (164.9 Nm, p=0.036) were both weaker than CB stems. Wired-TS stems showed no significant difference from non-wired-TS stems. Additionally, the study could not conclude that calcar wiring is stronger than diaphyseal wiring. All TS fractures occurred at the mid-stem, simulating a B-type fracture, while the addition of the diaphyseal wire shifted the fracture location more distally in four of seven stems (p=0.0699). This biomechanical study supports the clinical evidence that CB stems have stronger resistance to torque than TS stems and may explain lower risk of periprosthetic fracture. The addition of calcar or diaphyseal wires to TS stems resulted in no significant changes in peak torque to fracture. In patients at high risk of periprosthetic fracture, CB cemented stems should be considered.

Orthopaedic ProceedingsVol. 106-B(SUPP_16)
Mayo Clinic in Arizona (US)
Openalex Percentile: Top 31%
Orthopaedic implants and arthroplasty
0.57
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.