Development of a 3 T Conduction-Cooled Animal MRI Superconducting Magnet With a 300 mm Warm Bore

A 3 T conduction-cooled animal MRI superconducting magnet was designed, fabricated, and tested. The MRI magnet has a large clear warm bore of Ø300 mm and a center magnetic field strength of 3 T with the measured peak-to-peak homogeneity of 217 ppm in a 180 mm diameter sphere. The magnet consists of 8 coaxial coils placed symmetrically along the axial direction. All coils made with NbTi/Cu superconducting wires were connected in series and were operated with a current of 110 A with a stored energy of about 0.62 MJ. Passive protection based on a subdivided scheme and heater network design is adopted to protect the magnet from damage caused by a quench. A two-stage GM cryocooler with a second-stage cooling power of 1.8 W is used to cool the 3 T animal MRI magnet from room temperature to 4.2 K. In this paper, magnetic design, quench protection, fabrication, and test are presented.

Authors

Institutions

Publication Details

Journal
IEEE Transactions on Applied Superconductivity
Published
2024-08-19
DOI
https://doi.org/10.1109/tasc.2024.3420331
Citations
3
Primary Topic
Superconducting Materials and Applications
Type
article
Field-Weighted Citation Impact
0.55

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of a 3 T Conduction-Cooled Animal MRI Superconducting Magnet With a 300 mm Warm Bore

Qiuliang Wang, Shunzhong Chen, Wanshuo Sun, Huang Xie
3 citations
IEEE Transactions on Applied Superconductivity
Superconducting Materials and Applications
0.55
article

Development of a 3 T Conduction-Cooled Animal MRI Superconducting Magnet With a 300 mm Warm Bore

Qiuliang Wang, Shunzhong Chen, Wanshuo Sun, Huang Xie
article en
3 citations

Abstract

A 3 T conduction-cooled animal MRI superconducting magnet was designed, fabricated, and tested. The MRI magnet has a large clear warm bore of Ø300 mm and a center magnetic field strength of 3 T with the measured peak-to-peak homogeneity of 217 ppm in a 180 mm diameter sphere. The magnet consists of 8 coaxial coils placed symmetrically along the axial direction. All coils made with NbTi/Cu superconducting wires were connected in series and were operated with a current of 110 A with a stored energy of about 0.62 MJ. Passive protection based on a subdivided scheme and heater network design is adopted to protect the magnet from damage caused by a quench. A two-stage GM cryocooler with a second-stage cooling power of 1.8 W is used to cool the 3 T animal MRI magnet from room temperature to 4.2 K. In this paper, magnetic design, quench protection, fabrication, and test are presented.

IEEE Transactions on Applied SuperconductivityVol. 34(8)
Chinese Academy of Sciences (CN), Institute of Electrical Engineering (CN)
National Natural Science Foundation of China
Openalex Percentile: Top 38%
Superconducting Materials and Applications
0.55
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Development of a 3 T Conduction-Cooled Animal MRI Superconducting Magnet With a 300 mm Warm Bore — Qiuliang Wang, Shunzhong Chen, et al. · IEEE Transactions on Applied Superconductivity (2024) | TGRS Research Map | TGRS