Detection of Hemoglobin Constant Spring: A Comparison of Capillary Electrophoresis Versus High-Performance Liquid Chromatography

Hb CS was confirmed in all samples by molecular studies, revealing 92.3% heterozygous, 7.2% compound heterozygous, and 0.5% homozygous cases. CE detected 92.3% of heterozygous Hb CS cases, while HPLC detected only 48.2%. For compound heterozygous Hb CS, CE detected 100%, whereas HPLC detected 89.3%. Both homozygous cases were detected by CE and HPLC. The Pearson correlation coefficient test showed a significant linear relationship (p<0.001) between CE's zone 2 peak values and HPLC's C-window peaks (n=389). Conclusion: These findings highlight the efficacy of CE as a reliable method for Hb CS detection, suggesting its potential superiority over traditional HPLC in clinical settings.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2024-08-19
DOI
https://doi.org/10.7759/cureus.67228
Primary Topic
Hemoglobinopathies and Related Disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Detection of Hemoglobin Constant Spring: A Comparison of Capillary Electrophoresis Versus High-Performance Liquid Chromatography

Sumaiyah Adzahar, Adibah Daud, Mohammad Hudzaifah Nordin, Sharifatul Fatma Embong
Cureus
Hemoglobinopathies and Related Disorders
article

Detection of Hemoglobin Constant Spring: A Comparison of Capillary Electrophoresis Versus High-Performance Liquid Chromatography

Sumaiyah Adzahar, Adibah Daud, Mohammad Hudzaifah Nordin, Sharifatul Fatma Embong
article en

Abstract

Hb CS was confirmed in all samples by molecular studies, revealing 92.3% heterozygous, 7.2% compound heterozygous, and 0.5% homozygous cases. CE detected 92.3% of heterozygous Hb CS cases, while HPLC detected only 48.2%. For compound heterozygous Hb CS, CE detected 100%, whereas HPLC detected 89.3%. Both homozygous cases were detected by CE and HPLC. The Pearson correlation coefficient test showed a significant linear relationship (p<0.001) between CE's zone 2 peak values and HPLC's C-window peaks (n=389). Conclusion: These findings highlight the efficacy of CE as a reliable method for Hb CS detection, suggesting its potential superiority over traditional HPLC in clinical settings.

CureusVol. 16(8)
Sultan Zainal Abidin University (MY)
Openalex Percentile: Top 78%
Hemoglobinopathies and Related Disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.