Improving the efficiency of soft electron ionization by using the oscillatory movement of electrons

In process analytical chemistry, mass spectrometry analysis using a soft electron ionization (EI) source has qualitative advantages. However, the relatively small ionization cross-section of soft EI leads to lower sensitivity. To address this issue, a novel method has been developed to enhance the sensitivity of soft EI by utilizing a dual electron repeller and an ionization chamber to form a U-shaped electric field, causing electrons to oscillate within the field and effectively increasing the electron collision cross-sectional area. By combining with an electron lens, the virtual cathode effect at low electron energy can be reduced or even eliminated, thereby improving ionization efficiency. This method has resulted in a significant increase in signal intensity for m/z 18(H2O), with a factor of 4.2 at an electron energy of 25 eV and a factor of 3.75 at 20 eV, compared to the electron receiving mode. Additionally, it reduces the required emission current, which is beneficial for prolonging the life of the filament. The proposed technique is expected to expand the application of soft EI, particularly for rapid online analysis in process analytical chemistry such as catalyst research and chemical reaction process monitoring.

Authors

Institutions

Publication Details

Journal
Metrology and Measurement Systems
Published
2024-08-19
DOI
https://doi.org/10.24425/mms.2024.149696
Primary Topic
Mass Spectrometry Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Improving the efficiency of soft electron ionization by using the oscillatory movement of electrons

Xiang Fang, Zejian Huang, Mingfei Zhou, You Jiang et al.
Metrology and Measurement Systems
Mass Spectrometry Techniques and Applications
article

Improving the efficiency of soft electron ionization by using the oscillatory movement of electrons

Xiang Fang, Zejian Huang, Mingfei Zhou, You Jiang, Xinhua Dai
article en

Abstract

In process analytical chemistry, mass spectrometry analysis using a soft electron ionization (EI) source has qualitative advantages. However, the relatively small ionization cross-section of soft EI leads to lower sensitivity. To address this issue, a novel method has been developed to enhance the sensitivity of soft EI by utilizing a dual electron repeller and an ionization chamber to form a U-shaped electric field, causing electrons to oscillate within the field and effectively increasing the electron collision cross-sectional area. By combining with an electron lens, the virtual cathode effect at low electron energy can be reduced or even eliminated, thereby improving ionization efficiency. This method has resulted in a significant increase in signal intensity for m/z 18(H2O), with a factor of 4.2 at an electron energy of 25 eV and a factor of 3.75 at 20 eV, compared to the electron receiving mode. Additionally, it reduces the required emission current, which is beneficial for prolonging the life of the filament. The proposed technique is expected to expand the application of soft EI, particularly for rapid online analysis in process analytical chemistry such as catalyst research and chemical reaction process monitoring.

Metrology and Measurement Systems
Fudan University (CN), National Institute of Metrology (CN)
Affordable and clean energy
Openalex Percentile: Top 90%
Mass Spectrometry Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Improving the efficiency of soft electron ionization by using the oscillatory movement of electrons — Xiang Fang, Zejian Huang, et al. · Metrology and Measurement Systems (2024) | TGRS Research Map | TGRS