Potassium Ion-Selective Electrodes with BME-44 Ionophores Covalently Attached to Condensation-Cured Silicone Membranes

For ion-selective electrodes (ISEs) to be employed in wearable and implantable applications, the ion-selective membrane components should be biocompatible, and leaching of components, such as plasticizer or ionophore, out of the sensing membrane should be inhibited. To achieve this, we employed a plasticizer-free silicone as the membrane matrix and synthesized as the ionophore a derivative of the bis-crown ether based potassium ionophore BME-44, incorporating a triethoxysilyl functional group that covalently attaches to condensation-cured silicones during the curing process. Soxhlet extraction of these membranes with dichloromethane shows that up to 96% of the ionophore is attached to the silicone membrane during curing. We found that the covalently attachable BME-44 derivative can inadvertently adsorb onto high surface area carbon solid contacts before attaching to the silicone matrix if the curing of the silicone is performed in the presence of the high surface area carbon, resulting in depletion of ionophore from the membrane and yielding solid-contact ISEs with poor selectivity. In contrast, we observed Nernstian responses to K<sup>+</sup> in plasticizer-free silicone-based K<sup>+</sup> ISMs with either mobile BME-44 or the covalently attachable BME-44 derivative when the membranes were prepared on octane-thiol coated gold electrodes, where ionophore adsorption does not occur to a noticeable extent. As compared with ISMs doped with the mobile BME-44, ISMs prepared with the covalently attachable BME-44 derivative have better selectivity for K<sup>+</sup> vs Na<sup>+</sup> (<mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:mi>log</mml:mi><mml:mo>⁡</mml:mo><mml:msub><mml:mi>K</mml:mi><mml:mrow><mml:msup><mml:mi>K</mml:mi><mml:mo>+</mml:mo></mml:msup><mml:mo>,</mml:mo><mml:mi>N</mml:mi><mml:msup><mml:mi>a</mml:mi><mml:mo>+</mml:mo></mml:msup></mml:mrow></mml:msub></mml:math> values of -3.54 and <- 4.05 for mobile and covalently attachable BME-44, respectively) and lower resistance. This can be explained by a more homogeneous incorporation of the covalently attachable BME-44 derivative into the silicone matrix than is the case for the mobile BME-44.

Authors

Institutions

Publication Details

Journal
Langmuir
Published
2024-08-16
DOI
https://doi.org/10.1021/acs.langmuir.4c01726
Citations
1
Primary Topic
Analytical Chemistry and Sensors
Type
article
Field-Weighted Citation Impact
0.29

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Potassium Ion-Selective Electrodes with BME-44 Ionophores Covalently Attached to Condensation-Cured Silicone Membranes

Philippe Bühlmann, Xin Chen, Andreas Stein, Brian D. Spindler et al.
1 citations
Langmuir
Analytical Chemistry and Sensors
0.29
article

Potassium Ion-Selective Electrodes with BME-44 Ionophores Covalently Attached to Condensation-Cured Silicone Membranes

Philippe Bühlmann, Xin Chen, Andreas Stein, Brian D. Spindler, K. Graf
article en
1 citations

Abstract

For ion-selective electrodes (ISEs) to be employed in wearable and implantable applications, the ion-selective membrane components should be biocompatible, and leaching of components, such as plasticizer or ionophore, out of the sensing membrane should be inhibited. To achieve this, we employed a plasticizer-free silicone as the membrane matrix and synthesized as the ionophore a derivative of the bis-crown ether based potassium ionophore BME-44, incorporating a triethoxysilyl functional group that covalently attaches to condensation-cured silicones during the curing process. Soxhlet extraction of these membranes with dichloromethane shows that up to 96% of the ionophore is attached to the silicone membrane during curing. We found that the covalently attachable BME-44 derivative can inadvertently adsorb onto high surface area carbon solid contacts before attaching to the silicone matrix if the curing of the silicone is performed in the presence of the high surface area carbon, resulting in depletion of ionophore from the membrane and yielding solid-contact ISEs with poor selectivity. In contrast, we observed Nernstian responses to K<sup>+</sup> in plasticizer-free silicone-based K<sup>+</sup> ISMs with either mobile BME-44 or the covalently attachable BME-44 derivative when the membranes were prepared on octane-thiol coated gold electrodes, where ionophore adsorption does not occur to a noticeable extent. As compared with ISMs doped with the mobile BME-44, ISMs prepared with the covalently attachable BME-44 derivative have better selectivity for K<sup>+</sup> vs Na<sup>+</sup> (<mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:mi>log</mml:mi><mml:mo>⁡</mml:mo><mml:msub><mml:mi>K</mml:mi><mml:mrow><mml:msup><mml:mi>K</mml:mi><mml:mo>+</mml:mo></mml:msup><mml:mo>,</mml:mo><mml:mi>N</mml:mi><mml:msup><mml:mi>a</mml:mi><mml:mo>+</mml:mo></mml:msup></mml:mrow></mml:msub></mml:math> values of -3.54 and <- 4.05 for mobile and covalently attachable BME-44, respectively) and lower resistance. This can be explained by a more homogeneous incorporation of the covalently attachable BME-44 derivative into the silicone matrix than is the case for the mobile BME-44.

Langmuir
University of Minnesota (US)
Medtronic
Openalex Percentile: Top 50%
Analytical Chemistry and Sensors
0.29
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.