Design of automatic and integrated calibration device for dynamic characteristics of eddy-current sensor

Abstract To realize the dynamic calibration of eddy current sensor with high precision and high efficiency in a wide frequency range, a set of standard shaking table was designed as the actuator for the dynamic calibration of eddy current sensor, and the modal analysis and harmonic response analysis of the vibration disk were completed. At the same time, the control system was designed based on FPGA. The design of control system includes the design of signal generator with adjustable frequency and amplitude, data collector and communication module. Finally, the dynamic test and analysis of the eddy current sensor dynamic calibration disk was completed and the calibration accuracy of the standard vibration table is verified by comparative method. The deviation is less than 3% at the whole frequency band from 10-500HZ, which meets the design requirements.

Authors

Institutions

Publication Details

Journal
Journal of Physics Conference Series
Published
2024-08-01
DOI
https://doi.org/10.1088/1742-6596/2825/1/012015
Primary Topic
Non-Destructive Testing Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design of automatic and integrated calibration device for dynamic characteristics of eddy-current sensor

Xiang Xu, Jia Yao, Guanghua Deng, Shiqiang Qiu et al.
Journal of Physics Conference Series
Non-Destructive Testing Techniques
article

Design of automatic and integrated calibration device for dynamic characteristics of eddy-current sensor

Xiang Xu, Jia Yao, Guanghua Deng, Shiqiang Qiu, Jiang Ma
article en

Abstract

Abstract To realize the dynamic calibration of eddy current sensor with high precision and high efficiency in a wide frequency range, a set of standard shaking table was designed as the actuator for the dynamic calibration of eddy current sensor, and the modal analysis and harmonic response analysis of the vibration disk were completed. At the same time, the control system was designed based on FPGA. The design of control system includes the design of signal generator with adjustable frequency and amplitude, data collector and communication module. Finally, the dynamic test and analysis of the eddy current sensor dynamic calibration disk was completed and the calibration accuracy of the standard vibration table is verified by comparative method. The deviation is less than 3% at the whole frequency band from 10-500HZ, which meets the design requirements.

Journal of Physics Conference SeriesVol. 2825(1)
China Tobacco (CN)
Affordable and clean energy
Openalex Percentile: Top 87%
Non-Destructive Testing Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.