Investigation of the beam quality and dose rate dependence of PAKAG polymer gel dosimeter in optical readout technique

Abstract This study aims to evaluate the optical response dependence of the PAKAG polymer gel dosimeter on photon energy and dose rate. The produced gel dosimeters were irradiated using a Varian CL 21EX medical linear accelerator with delivered doses of 0, 2, 4, 6, 8, and 10 Gy. To examine the response dependence on the delivered dose rate, dose rates of 50, 100, 200, and 350 cGy min −1 were investigated. Additionally, two incident beam qualities of 6 and 18 MV were examined to study the response dependence on the incident beam energy. The irradiated polymer gel dosimeters were readout using a UV–vis spectrophotometer in the 300 to 800 nm scan range. The results reveal that a wide variation in dose rate (50–350 cGy.min −1 ) influences the absorbance-dose response and the sensitivity of PAKAG gel. However, smaller variations did not show a significant effect on the response. Furthermore, the response changed insignificantly with beam quality for investigated energies. It was concluded that the optical reading response of the PAKAG polymer gel dosimeter is satisfactorily independent of external parameters, including dose rate and incident beam quality.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2024-08-16
DOI
https://doi.org/10.1088/2057-1976/ad7032
Citations
2
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
1.22

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Investigation of the beam quality and dose rate dependence of PAKAG polymer gel dosimeter in optical readout technique

Seyed Mohammad Mahdi Abtahi, Fatemeh Habibi
2 citations
Biomedical Physics & Engineering Express
Advanced Radiotherapy Techniques
1.22
article

Investigation of the beam quality and dose rate dependence of PAKAG polymer gel dosimeter in optical readout technique

Seyed Mohammad Mahdi Abtahi, Fatemeh Habibi
article en
2 citations

Abstract

Abstract This study aims to evaluate the optical response dependence of the PAKAG polymer gel dosimeter on photon energy and dose rate. The produced gel dosimeters were irradiated using a Varian CL 21EX medical linear accelerator with delivered doses of 0, 2, 4, 6, 8, and 10 Gy. To examine the response dependence on the delivered dose rate, dose rates of 50, 100, 200, and 350 cGy min −1 were investigated. Additionally, two incident beam qualities of 6 and 18 MV were examined to study the response dependence on the incident beam energy. The irradiated polymer gel dosimeters were readout using a UV–vis spectrophotometer in the 300 to 800 nm scan range. The results reveal that a wide variation in dose rate (50–350 cGy.min −1 ) influences the absorbance-dose response and the sensitivity of PAKAG gel. However, smaller variations did not show a significant effect on the response. Furthermore, the response changed insignificantly with beam quality for investigated energies. It was concluded that the optical reading response of the PAKAG polymer gel dosimeter is satisfactorily independent of external parameters, including dose rate and incident beam quality.

Biomedical Physics & Engineering ExpressVol. 10(5)
Imam Khomeini International University (IR)
Imam Khomeini International University
Affordable and clean energy
Openalex Percentile: Top 22%
Advanced Radiotherapy Techniques
1.22
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Investigation of the beam quality and dose rate dependence of PAKAG polymer gel dosimeter in optical readout technique — Seyed Mohammad Mahdi Abtahi, Fatemeh Habibi · Biomedical Physics & Engineering Express (2024) | TGRS Research Map | TGRS