Evaluation of a Slim Modiolar Electrode Array: A Temporal Bone Study
TB evaluation revealed that the SM electrode exerts minimal insertion forces on cochlear structures, produces no histologic evidence of trauma, and reliably assumes the perimodiolar position. Nonstandard cochleostomy location, improper sheath insertion depth, or premature deployment of the electrode may lead to suboptimal outcomes.
Authors
- Claudiu Treaba
- Leena Asfour (ORCID: https://orcid.org/0000-0003-0926-3803)
- Jonathon R. Kirk (ORCID: https://orcid.org/0000-0003-4945-4399)
- Frank Risi
- J. Roland Thomas
Institutions
- New York University (US)
Publication Details
- Journal
- Otology & Neurotology
- Published
- 2024-08-15
- DOI
- https://doi.org/10.1097/mao.0000000000004253
- Citations
- 1
- Primary Topic
- Hearing Loss and Rehabilitation
- Type
- article
- Field-Weighted Citation Impact
- 0.34