Intraventricular flow dynamics and thrombosis risk with a left ventricular assist device using accelerated thrombosis modeling and stress-blended eddy simulation

Thrombosis remains a significant complication in blood-contacting medical devices. Computational studies on flow dynamics within a left ventricular assist device (LVAD) supported left ventricle (LV) have been conducted, yet there is a lack of detailed models capable of comprehensive LV flow dynamics analysis to predict localized thrombus risk. This study aimed to address these gaps by employing a scale-resolving turbulence model combined with an accelerated thrombosis model for intraventricular flow disturbed by an LVAD. A patient-specific LV model of a post-VAD patient was implanted with a HeartMate3 cannula. A complete heart failure scenario with a 5 L/min LVAD outflow and rigid walls was developed. The blood turbulence field was modeled using a Stress-blended eddy simulation turbulence model. The accelerated thrombosis model involved three convection-diffusion equations to trace the non-activated platelets, activated platelets, and adenosine diphosphate. Results were compared with residence time, a simplified thrombosis index commonly used in the literature. Blood residence time peaked at 7.5 s along the cannula surface and exceeded 3.5 s around the aortic root and behind the mitral leaflets. Concurrently, the scaled activated platelet concentration reached its maximum value on the cannula surface toward the cannula outlet. The activation rates of platelets indicate a negligible activation, suggesting that thrombosis occurring in the LV of LVAD patients is primarily due to prolonged blood residence, leading to increased coagulation.

Authors

Institutions

Publication Details

Journal
Physics of Fluids
Published
2024-08-01
DOI
https://doi.org/10.1063/5.0214303
Citations
7
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
1.29

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Intraventricular flow dynamics and thrombosis risk with a left ventricular assist device using accelerated thrombosis modeling and stress-blended eddy simulation

Shaun D. Gregory, Mehrdad Khamooshi, Marjan Azimi
7 citations
Physics of Fluids
Mechanical Circulatory Support Devices
1.29
article

Intraventricular flow dynamics and thrombosis risk with a left ventricular assist device using accelerated thrombosis modeling and stress-blended eddy simulation

Shaun D. Gregory, Mehrdad Khamooshi, Marjan Azimi
article en
7 citations

Abstract

Thrombosis remains a significant complication in blood-contacting medical devices. Computational studies on flow dynamics within a left ventricular assist device (LVAD) supported left ventricle (LV) have been conducted, yet there is a lack of detailed models capable of comprehensive LV flow dynamics analysis to predict localized thrombus risk. This study aimed to address these gaps by employing a scale-resolving turbulence model combined with an accelerated thrombosis model for intraventricular flow disturbed by an LVAD. A patient-specific LV model of a post-VAD patient was implanted with a HeartMate3 cannula. A complete heart failure scenario with a 5 L/min LVAD outflow and rigid walls was developed. The blood turbulence field was modeled using a Stress-blended eddy simulation turbulence model. The accelerated thrombosis model involved three convection-diffusion equations to trace the non-activated platelets, activated platelets, and adenosine diphosphate. Results were compared with residence time, a simplified thrombosis index commonly used in the literature. Blood residence time peaked at 7.5 s along the cannula surface and exceeded 3.5 s around the aortic root and behind the mitral leaflets. Concurrently, the scaled activated platelet concentration reached its maximum value on the cannula surface toward the cannula outlet. The activation rates of platelets indicate a negligible activation, suggesting that thrombosis occurring in the LV of LVAD patients is primarily due to prolonged blood residence, leading to increased coagulation.

Physics of FluidsVol. 36(8)
Queensland University of Technology (AU), Monash University (AU)
National Heart Foundation of Australia, National Health and Medical Research Council
Good health and well-being
Openalex Percentile: Top 22%
Mechanical Circulatory Support Devices
1.29
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.