Characteristics of 2N4416 Si JFETs for radiation spectrometer charge sensitive preamplifiers

The input transistors (JFETs) of feedback resistor-less charge sensitive preamplifiers (CSPs), for photon and charged particle counting radiation spectrometers, that use semiconductors as their detectors, are operated in an unusual forward biased mode; consequently, manufacturer datasheets cannot necessarily be relied upon, as they do not specify the characteristics of the JFETs when they are operated with the gate-to-source junction forward biased. This experimental study, in which 165 2N4416 and 2N4416A NJ26 silicon junction field effect transistors (JFETs) were electrically characterised at ambient laboratory temperature, focused on the JFETs’ performances in the context of their use as input transistors of feedback resistor-less charge sensitive preamplifiers (CSPs). The experimental study found evidence which supports much of the common practice of the skilled CSP designer but which had not been documented with results in the open literature until now. This includes, inter alia, that: the 2N4416 performs better than the 2N4416A; not all transistors even of the same type have comparable performance when operated in forward bias mode; only a small proportion of 2N4416 and 2N4416A transistors are of desirable performance characteristics for CSPs; and, individual screening of each candidate JFET is required if an optimally performing CSP is to be produced – even when using JFETs of the same type from the same batch. The methodology and analysis documented here will also be useful to researchers developing new wide bandgap semiconductor JFETs for high temperature tolerant and radiation-hard CSPs.

Authors

Institutions

Publication Details

Journal
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Published
2024-08-15
DOI
https://doi.org/10.1016/j.nima.2024.169745
Citations
2
Primary Topic
Particle Detector Development and Performance
Type
article
Field-Weighted Citation Impact
1.32

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Characteristics of 2N4416 Si JFETs for radiation spectrometer charge sensitive preamplifiers

M. D. C. Whitaker, C.S. Bodie, A.M. Barnett
2 citations
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Particle Detector Development and Performance
1.32
article

Characteristics of 2N4416 Si JFETs for radiation spectrometer charge sensitive preamplifiers

M. D. C. Whitaker, C.S. Bodie, A.M. Barnett
article en
2 citations

Abstract

The input transistors (JFETs) of feedback resistor-less charge sensitive preamplifiers (CSPs), for photon and charged particle counting radiation spectrometers, that use semiconductors as their detectors, are operated in an unusual forward biased mode; consequently, manufacturer datasheets cannot necessarily be relied upon, as they do not specify the characteristics of the JFETs when they are operated with the gate-to-source junction forward biased. This experimental study, in which 165 2N4416 and 2N4416A NJ26 silicon junction field effect transistors (JFETs) were electrically characterised at ambient laboratory temperature, focused on the JFETs’ performances in the context of their use as input transistors of feedback resistor-less charge sensitive preamplifiers (CSPs). The experimental study found evidence which supports much of the common practice of the skilled CSP designer but which had not been documented with results in the open literature until now. This includes, inter alia, that: the 2N4416 performs better than the 2N4416A; not all transistors even of the same type have comparable performance when operated in forward bias mode; only a small proportion of 2N4416 and 2N4416A transistors are of desirable performance characteristics for CSPs; and, individual screening of each candidate JFET is required if an optimally performing CSP is to be produced – even when using JFETs of the same type from the same batch. The methodology and analysis documented here will also be useful to researchers developing new wide bandgap semiconductor JFETs for high temperature tolerant and radiation-hard CSPs.

Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated EquipmentVol. 1068
University of Sussex (GB), University of Warwick (GB)
Leverhulme Trust, Science and Technology Facilities Council
Affordable and clean energy
Openalex Percentile: Top 11%
Particle Detector Development and Performance
1.32
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Characteristics of 2N4416 Si JFETs for radiation spectrometer charge sensitive preamplifiers — M. D. C. Whitaker, C.S. Bodie, et al. · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment (2024) | TGRS Research Map | TGRS