Scatter estimation and correction using time-of-flight and deconvolution in x-ray medical imaging

<i>Objective.</i>Time-of-flight (TOF) scatter rejection requires a total timing jitter, including the detector timing jitter and the x-ray source's pulses width, of 50 ps or less to mitigate most of the effects of scattered photons in radiography and CT imaging. However, since the total contribution of the source and detector to the timing jitter can be retrieved during an acquisition with nothing between the source and detector, it can be demonstrated that this contribution may be partially removed to improve the image quality.<i>Approach.</i>A scatter correction method using iterative deconvolution of the measured time point-spread function estimates the number of scattered photons detected in each pixel. To evaluate the quality of the estimation, GATE was used to simulate the radiography of a water cylinder with bone inserts, and a head and torso in a system with total timing jitters from 100 ps up to 500 ps full-width-at-half-maximum (FWHM).<i>Main results.</i>With a total timing jitter of 200 ps FWHM, 89% of the contrast degradation caused by scattered photons was recovered in a head and torso radiography, compared to 28% with a simple time threshold method. Corrected images using the estimation have a percent root-mean square error between 2% and 14% in both phantoms with timing jitters from 100 to 500 ps FWHM which is lower than the error achieved with scatter rejection alone at 100 ps FWHM.<i>Significance.</i>TOF x-ray imaging has the potential to mitigate the effects of the scattering contribution and offers an alternative to anti-scatter grids that avoids loss of primary photons. Compare to simple TOF scatter rejection using only a threshold, the deconvolution estimation approach has lower requirements on both the source and detector. These requirements are now within reach of state-of-the-art systems.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2024-08-15
DOI
https://doi.org/10.1088/1361-6560/ad700e
Citations
5
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
2.02

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Scatter estimation and correction using time-of-flight and deconvolution in x-ray medical imaging

Julien Rossignol, Yves Bérubé-Lauzière, Réjean Fontaine, Audrey Corbeil Therrien et al.
5 citations
Physics in Medicine and Biology
Radiation Dose and Imaging
2.02
article

Scatter estimation and correction using time-of-flight and deconvolution in x-ray medical imaging

Julien Rossignol, Yves Bérubé-Lauzière, Réjean Fontaine, Audrey Corbeil Therrien, Maëlle Fromont, Gabriel Bélanger
article en
5 citations

Abstract

<i>Objective.</i>Time-of-flight (TOF) scatter rejection requires a total timing jitter, including the detector timing jitter and the x-ray source's pulses width, of 50 ps or less to mitigate most of the effects of scattered photons in radiography and CT imaging. However, since the total contribution of the source and detector to the timing jitter can be retrieved during an acquisition with nothing between the source and detector, it can be demonstrated that this contribution may be partially removed to improve the image quality.<i>Approach.</i>A scatter correction method using iterative deconvolution of the measured time point-spread function estimates the number of scattered photons detected in each pixel. To evaluate the quality of the estimation, GATE was used to simulate the radiography of a water cylinder with bone inserts, and a head and torso in a system with total timing jitters from 100 ps up to 500 ps full-width-at-half-maximum (FWHM).<i>Main results.</i>With a total timing jitter of 200 ps FWHM, 89% of the contrast degradation caused by scattered photons was recovered in a head and torso radiography, compared to 28% with a simple time threshold method. Corrected images using the estimation have a percent root-mean square error between 2% and 14% in both phantoms with timing jitters from 100 to 500 ps FWHM which is lower than the error achieved with scatter rejection alone at 100 ps FWHM.<i>Significance.</i>TOF x-ray imaging has the potential to mitigate the effects of the scattering contribution and offers an alternative to anti-scatter grids that avoids loss of primary photons. Compare to simple TOF scatter rejection using only a threshold, the deconvolution estimation approach has lower requirements on both the source and detector. These requirements are now within reach of state-of-the-art systems.

Physics in Medicine and BiologyVol. 69(17)
Université de Sherbrooke (CA)
Natural Sciences and Engineering Research Council of Canada
Openalex Percentile: Top 13%
Radiation Dose and Imaging
2.02
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.