The influence of calibration on bias in quality control and patient results for TSH on Vitros XT 7600 analyzer

Even though the relative bias exceeded allowable criteria for the QC samples, this was not manifested on the patients' samples.

Authors

Institutions

Publication Details

Journal
Biochemia Medica
Published
2024-08-05
DOI
https://doi.org/10.11613/bm.2024.030703
Primary Topic
Thyroid Disorders and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The influence of calibration on bias in quality control and patient results for TSH on Vitros XT 7600 analyzer

Josko Ivica, Jill Boreyko
Biochemia Medica
Thyroid Disorders and Treatments
article

The influence of calibration on bias in quality control and patient results for TSH on Vitros XT 7600 analyzer

Josko Ivica, Jill Boreyko
article en

Abstract

Even though the relative bias exceeded allowable criteria for the QC samples, this was not manifested on the patients' samples.

Biochemia MedicaVol. 34(3)
Hamilton Regional Laboratory Medicine Program (CA), McMaster University (CA)
Openalex Percentile: Top 79%
Thyroid Disorders and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

The influence of calibration on bias in quality control and patient results for TSH on Vitros XT 7600 analyzer — Josko Ivica, Jill Boreyko · Biochemia Medica (2024) | TGRS Research Map | TGRS