D-dimer assay interference detected by the discrepancy in D-dimer concentrations at different dilutions: a case report

This case report describes interference from heterophilic antibodies in D-dimer assay. The interference was suspected due to discrepancies between D-dimer concentrations in the original sample and diluted samples, as well as inconsistent clinical findings. The patient's medical history, laboratory results, and imaging studies were considered in the investigation. Heterophilic antibodies, likely developed during the SARS-CoV-2 infection, were identified as the probable cause of interference. The interference was confirmed through various methods, including dilution studies, blocking heterophilic antibodies, and comparing results with an alternative D-dimer method. This case highlights the importance of recognizing and addressing interference in D-dimer testing, emphasizing the need for collaboration between clinicians and laboratory specialists.

Authors

Institutions

Publication Details

Journal
Biochemia Medica
Published
2024-08-05
DOI
https://doi.org/10.11613/bm.2024.031001
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

D-dimer assay interference detected by the discrepancy in D-dimer concentrations at different dilutions: a case report

Vikica Buljanović, Iva Lukić, Vatroslav Šerić, Marija Milić et al.
Biochemia Medica
Clinical Laboratory Practices and Quality Control
article

D-dimer assay interference detected by the discrepancy in D-dimer concentrations at different dilutions: a case report

Vikica Buljanović, Iva Lukić, Vatroslav Šerić, Marija Milić, Mirna Kirin, Dejana Brkić Barbarić
article en

Abstract

This case report describes interference from heterophilic antibodies in D-dimer assay. The interference was suspected due to discrepancies between D-dimer concentrations in the original sample and diluted samples, as well as inconsistent clinical findings. The patient's medical history, laboratory results, and imaging studies were considered in the investigation. Heterophilic antibodies, likely developed during the SARS-CoV-2 infection, were identified as the probable cause of interference. The interference was confirmed through various methods, including dilution studies, blocking heterophilic antibodies, and comparing results with an alternative D-dimer method. This case highlights the importance of recognizing and addressing interference in D-dimer testing, emphasizing the need for collaboration between clinicians and laboratory specialists.

Biochemia MedicaVol. 34(3)
Croatian Institute of Transfusion Medicine (HR), University of Osijek (HR)
Openalex Percentile: Top 82%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

D-dimer assay interference detected by the discrepancy in D-dimer concentrations at different dilutions: a case report — Vikica Buljanović, Iva Lukić, et al. · Biochemia Medica (2024) | TGRS Research Map | TGRS