PET Prototype Based on Scintillation Detectors GAGG-SiPM Coupled to 32-Channel Petiroc2A Chip
We present an early-stage prototype of a positron emission tomograph (PET) based on 3 × 3 × 20 mm GAGG(Ce) scintillators optically coupled to 3 × 3 mm silicon photomultipliers (SiPMs) (Onsemi fc30035-smt). The detectors are read out by a Weeroc 32-channel Petiroc2A chip, allowing for precise charge and time measurements and designed specifically for SiPMs readout. In the report, the energy and time resolutions of the system measured with Ti-44 source are presented, the imaging capabilities of the system are discussed.
Authors
- Filipp Dubinin (ORCID: https://orcid.org/0000-0003-2182-2727)
- G. Dolganov (ORCID: https://orcid.org/0000-0002-6394-9219)
- А. Д. Конотоп (ORCID: https://orcid.org/0009-0008-3999-3210)
- A. Zhurkina
- N. I. Boyko
Institutions
- P.N. Lebedev Physical Institute of the Russian Academy of Sciences (RU)
- Kurchatov Institute (RU)
- National Research Nuclear University MEPhI (RU)
Publication Details
- Journal
- Physics of Particles and Nuclei Letters
- Published
- 2024-08-01
- DOI
- https://doi.org/10.1134/s1547477124701164
- Primary Topic
- Medical Imaging Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00