Multibraided Fixed Retainers with Different Diameters after Magnetic Resonance Imaging (MRI): In Vitro Study Investigating Temperature Changes and Bonding Efficacy

Objectives: Orthodontists are often asked to remove fixed retainers before patients undergo magnetic resonance imaging (MRI). The present in vitro study was designed to analyze the heating and bonding efficacy of stainless steel multibraided fixed retainers after 1.5- and 3-tesla (T) MRI. Materials and methods: A total of 180 human mandibular incisors were used to create 45 specimens of four teeth each, divided into nine groups. Handmade multibraided fixed retainers of three different sizes, defined by the diameter of the initial wire used (0.008″, 0.010″ and 0.012″), were tested. Three groups underwent MRI at 1.5 T, another three groups underwent MRI at 3 T and the last three groups did not undergo MRI. Temperature was assessed before and after MRI. Shear bond strength (SBS) and adhesive remnant index (ARI) were assessed after MRI for all groups. Data were statistically analyzed (p < 0.05). Results: After 1.5 T exposure, no significant temperature increase from T0 to T1 was observed in any of the groups (p > 0.05). Regarding the 3 T groups, a significant difference from T0 to T1 was found for all the groups (p < 0.05). Temperature changes were not clinically relevant, as they were less than 1 °C for all groups except for group 3 (ΔT0–T1: 1.18 ± 0.3 °C) and group 6 (ΔT0–T1: 1.12 ± 0.37 °C). Furthermore, there were no significant differences between the temperature variations associated with different wire diameters (p > 0.05). Conclusions: No significant changes in SBS or ARI were found (p > 0.05). Clinical significance: Since overheating was irrelevant and adhesion values did not change, the tested devices were concluded to be safe for MRI examinations at 1.5 T and 3 T.

Authors

Institutions

Publication Details

Journal
Dentistry Journal
Published
2024-08-13
DOI
https://doi.org/10.3390/dj12080255
Primary Topic
Dental materials and restorations
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Multibraided Fixed Retainers with Different Diameters after Magnetic Resonance Imaging (MRI): In Vitro Study Investigating Temperature Changes and Bonding Efficacy

Andrea Scribante, Cinzia Rizzi, Lorenzo Preda, Karin Bertino et al.
Dentistry Journal
Dental materials and restorations
article

Multibraided Fixed Retainers with Different Diameters after Magnetic Resonance Imaging (MRI): In Vitro Study Investigating Temperature Changes and Bonding Efficacy

Andrea Scribante, Cinzia Rizzi, Lorenzo Preda, Karin Bertino, Maurizio Pascadopoli, Domenico Sfondrini, Maria Francesca Sfondrini, Paola Gandini
article en

Abstract

Objectives: Orthodontists are often asked to remove fixed retainers before patients undergo magnetic resonance imaging (MRI). The present in vitro study was designed to analyze the heating and bonding efficacy of stainless steel multibraided fixed retainers after 1.5- and 3-tesla (T) MRI. Materials and methods: A total of 180 human mandibular incisors were used to create 45 specimens of four teeth each, divided into nine groups. Handmade multibraided fixed retainers of three different sizes, defined by the diameter of the initial wire used (0.008″, 0.010″ and 0.012″), were tested. Three groups underwent MRI at 1.5 T, another three groups underwent MRI at 3 T and the last three groups did not undergo MRI. Temperature was assessed before and after MRI. Shear bond strength (SBS) and adhesive remnant index (ARI) were assessed after MRI for all groups. Data were statistically analyzed (p < 0.05). Results: After 1.5 T exposure, no significant temperature increase from T0 to T1 was observed in any of the groups (p > 0.05). Regarding the 3 T groups, a significant difference from T0 to T1 was found for all the groups (p < 0.05). Temperature changes were not clinically relevant, as they were less than 1 °C for all groups except for group 3 (ΔT0–T1: 1.18 ± 0.3 °C) and group 6 (ΔT0–T1: 1.12 ± 0.37 °C). Furthermore, there were no significant differences between the temperature variations associated with different wire diameters (p > 0.05). Conclusions: No significant changes in SBS or ARI were found (p > 0.05). Clinical significance: Since overheating was irrelevant and adhesion values did not change, the tested devices were concluded to be safe for MRI examinations at 1.5 T and 3 T.

Dentistry JournalVol. 12(8)
University of Pavia (IT), National Center for Oncological Hadrontherapy (IT), Policlinico San Matteo Fondazione (IT), Istituti di Ricovero e Cura a Carattere Scientifico (IT)
Openalex Percentile: Top 85%
Dental materials and restorations
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.