Timing offset calibration for TOF PET using stationary line source scans at multiple positions

<i>Background</i>. Accurate timing offset calibration is crucial for time-of-flight (TOF) positron emission tomography (PET) to mitigate image artifacts and improve quantitative accuracy. However, existing methods are often time-consuming, complex, or costly.<i>Objective</i>. This paper presents a method for TOF PET timing offset calibration that eliminates the need for costly equipment, phantoms, short-half-life sources, and precise source positioning.<i>Approach</i>. We estimate channel timing offsets using stationary scans of a<sup>68</sup>Ge line source, typically used for routine quality control, at a minimum of three non-coplanar positions, with each position scanned for two minutes. The line source positions are accurately determined by applying a simple algorithm to their reconstructed images, allowing precise calculation of arrival time differences. Channel timing offsets are estimated by solving a least squares problem. This method is assessed through analyses of phantoms and patient images using a RAYSOLUTION DigitMI 930 scanner.<i>Main results</i>. The estimated timing offsets ranged from -500 ps to 500 ps across all channels. Calibration with a minimum of three scanned positions was sufficient to correct these offsets, achieving less than a 1% discrepancy across various metrics of the image quality (IQ) phantom compared to 12 positions. This calibration significantly reduced edge artifacts in TOF reconstruction of both phantoms and patients. Furthermore, the IQ phantom displayed a 14% increase in average contrast recovery, a 61% reduction in average background variability across all spheres, and a 90% reduction in average residual error. Consistent with the phantom results, patient data revealed enhancements in maximum standardized uptake values (SUVmax) from 14% to 55% for lesions measuring 6 mm to 14 mm. The calibration also improved lesion-to-background contrast and eliminated artifacts caused by the spillover effect of the kidneys and bladder.<i>Significance</i>. The proposed method is fast, user-friendly, and cost-effective, effectively improving lesion detection and diagnostic accuracy.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2024-08-13
DOI
https://doi.org/10.1088/1361-6560/ad6edb
Citations
1
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.40

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Timing offset calibration for TOF PET using stationary line source scans at multiple positions

Junpeng Hu, Jie Zhang, Lei Fang, Peng Xiao et al.
1 citations
Physics in Medicine and Biology
Medical Imaging Techniques and Applications
0.40
article

Timing offset calibration for TOF PET using stationary line source scans at multiple positions

Junpeng Hu, Jie Zhang, Lei Fang, Peng Xiao, Xuan Zhang, Ang Li, Xiaoyun Zhou, Bingxuan Li, Fei Li, Qingguo Xie
article en
1 citations

Abstract

<i>Background</i>. Accurate timing offset calibration is crucial for time-of-flight (TOF) positron emission tomography (PET) to mitigate image artifacts and improve quantitative accuracy. However, existing methods are often time-consuming, complex, or costly.<i>Objective</i>. This paper presents a method for TOF PET timing offset calibration that eliminates the need for costly equipment, phantoms, short-half-life sources, and precise source positioning.<i>Approach</i>. We estimate channel timing offsets using stationary scans of a<sup>68</sup>Ge line source, typically used for routine quality control, at a minimum of three non-coplanar positions, with each position scanned for two minutes. The line source positions are accurately determined by applying a simple algorithm to their reconstructed images, allowing precise calculation of arrival time differences. Channel timing offsets are estimated by solving a least squares problem. This method is assessed through analyses of phantoms and patient images using a RAYSOLUTION DigitMI 930 scanner.<i>Main results</i>. The estimated timing offsets ranged from -500 ps to 500 ps across all channels. Calibration with a minimum of three scanned positions was sufficient to correct these offsets, achieving less than a 1% discrepancy across various metrics of the image quality (IQ) phantom compared to 12 positions. This calibration significantly reduced edge artifacts in TOF reconstruction of both phantoms and patients. Furthermore, the IQ phantom displayed a 14% increase in average contrast recovery, a 61% reduction in average background variability across all spheres, and a 90% reduction in average residual error. Consistent with the phantom results, patient data revealed enhancements in maximum standardized uptake values (SUVmax) from 14% to 55% for lesions measuring 6 mm to 14 mm. The calibration also improved lesion-to-background contrast and eliminated artifacts caused by the spillover effect of the kidneys and bladder.<i>Significance</i>. The proposed method is fast, user-friendly, and cost-effective, effectively improving lesion detection and diagnostic accuracy.

Physics in Medicine and BiologyVol. 69(17)
University of Science and Technology of China (CN), Anhui University (CN), Anhui Medical University (CN), Huazhong Agricultural University (CN), National Science Centre (PL), Second Affiliated Hospital of Anhui Medical University (CN), Huazhong University of Science and Technology (CN)
National Natural Science Foundation of China
Openalex Percentile: Top 36%
Medical Imaging Techniques and Applications
0.40
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.