Comparative analysis of anterior corneal surface characteristics measured by two topographers and two autokeratometers

Background: Precise quantification of corneal parameters enhances the clinical of the anterior corneal surface for various clinical assessments, with ongoing research focusing on optimizing everyday practice through device comparison studies. The study aimed to compare the agreement on the anterior corneal surface characteristics values measured by four commercially available devices: EyeSys 2000, Oculus Easygraph, ARK700A and ARK510A. Methods: Observational cross-sectional study involving 50 eyes from healthy volunteer recruited among subjects attending for an eye examination. Corneal radius minimum keratometry reading (min-K) and maximum keratometry reading (max-K), astigmatism axis and eccentricity were measured in all participants with EyeSys, Oculus Easygraph, ARK700A and ARK510A. Measurements were performed three times only in the right eye in random order by four optometrists who were unaware of the results of the other researchers. Results: A statistically significant difference occurred for the measurements of the corneal radius (Greenhouse-Geisser and Huynh-Feldt: P≤0.001 for both max-K and min-K radius) and the axis of astigmatism (Friedman: P<0.001), between the four devices. Pairwise comparisons revealed no radius differences between Oculus Easygraph and ARK510A (Sidak: min-K radius, P=0.231; max-K radius, P=0.224), and no axis differences in all pairwise analyses that have not imply the EyeSys (Wilcoxon: all P≥0.130). There was no statistical difference in the eccentricity value obtained by the topographers (paired t-test: P=0.507). Conclusions: The present study showed that EyeSys 2000 measurements differed from those of the other devices across all parameters. Alternatively, both topographers showed a perfect agreement concerning eccentricity measurements in everyday practice.

Authors

Institutions

Publication Details

Journal
AME Medical Journal
Published
2024-08-12
DOI
https://doi.org/10.21037/amj-23-257
Primary Topic
Corneal surgery and disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative analysis of anterior corneal surface characteristics measured by two topographers and two autokeratometers

Hugo Pena‐Verdeal, Eva Yebra‐Pimentel, Jacobo García-Queiruga, Noelia Nores-Palmas et al.
AME Medical Journal
Corneal surgery and disorders
article

Comparative analysis of anterior corneal surface characteristics measured by two topographers and two autokeratometers

Hugo Pena‐Verdeal, Eva Yebra‐Pimentel, Jacobo García-Queiruga, Noelia Nores-Palmas, Verónica Noya-Padin, Alba Castro-Giraldez
article en

Abstract

Background: Precise quantification of corneal parameters enhances the clinical of the anterior corneal surface for various clinical assessments, with ongoing research focusing on optimizing everyday practice through device comparison studies. The study aimed to compare the agreement on the anterior corneal surface characteristics values measured by four commercially available devices: EyeSys 2000, Oculus Easygraph, ARK700A and ARK510A. Methods: Observational cross-sectional study involving 50 eyes from healthy volunteer recruited among subjects attending for an eye examination. Corneal radius minimum keratometry reading (min-K) and maximum keratometry reading (max-K), astigmatism axis and eccentricity were measured in all participants with EyeSys, Oculus Easygraph, ARK700A and ARK510A. Measurements were performed three times only in the right eye in random order by four optometrists who were unaware of the results of the other researchers. Results: A statistically significant difference occurred for the measurements of the corneal radius (Greenhouse-Geisser and Huynh-Feldt: P≤0.001 for both max-K and min-K radius) and the axis of astigmatism (Friedman: P<0.001), between the four devices. Pairwise comparisons revealed no radius differences between Oculus Easygraph and ARK510A (Sidak: min-K radius, P=0.231; max-K radius, P=0.224), and no axis differences in all pairwise analyses that have not imply the EyeSys (Wilcoxon: all P≥0.130). There was no statistical difference in the eccentricity value obtained by the topographers (paired t-test: P=0.507). Conclusions: The present study showed that EyeSys 2000 measurements differed from those of the other devices across all parameters. Alternatively, both topographers showed a perfect agreement concerning eccentricity measurements in everyday practice.

AME Medical JournalVol. 10
Universidade de Santiago de Compostela (ES), Instituto de Investigación Sanitaria de Santiago (ES)
Openalex Percentile: Top 82%
Corneal surgery and disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.