Abstract To address the inadequacies of traditional seawater total alkalinity detection methods, which fail to meet the demands for continuous on-site monitoring at sea, a seawater total alkalinity analyzer based on single-point titration and spectrophotometry has been developed. This paper focuses on the design of the analyzer’s system architecture, hardware, and control systems. The instrument adopts a modular design concept and utilizes microfluidic technology for detection, with the STM32 serving as the control core. Through the design of the control system for the detection instrument, accurate in-situ automatic measurement of seawater total alkalinity is achieved. Laboratory testing and analysis demonstrate that the instrument’s measurement relative error is less than 1%, satisfying the accuracy requirements for seawater total alkalinity measurement. This research provides critical monitoring data support for a timely understanding of marine ecological environments and the marine carbon cycle.
Authors
- Jing Zeng (ORCID: https://orcid.org/0000-0001-9975-3130)
- Bingwei Wu
- Yingying Zhang (ORCID: https://orcid.org/0000-0001-9796-919X)
- Hongfan Wang
- Da Yuan
- Maotong Hao
Institutions
- Institute of Oceanographic Instrumentation (CN)
Publication Details
- Journal
- Journal of Physics Conference Series
- Published
- 2024-08-01
- DOI
- https://doi.org/10.1088/1742-6596/2815/1/012045
- Primary Topic
- Ocean Acidification Effects and Responses
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- National Natural Science Foundation of China
- Shandong Academy of Sciences
- Qilu University of Technology