Optimized Design of High-Temperature Resistant Air-Coupled Ultrasonic Transducer (ACUT) Based on Single Acoustic Matching Layer

A high-temperature resistant air-coupled ultrasonic transducer (ACUT) was optimally designed with a single acoustic matching layer structure. The matching layer was prepared with an organosilicon substrate instead of traditional epoxy resin mixed with hollow glass microspheres. Acoustic performance tests showed that the matching layer achieved an acoustic impedance of 1.06 MRayl and an acoustic attenuation coefficient of 0.47 dB/mm. The ACUT was clamped with a prestressed bolt structure instead of an adhesive layer and encapsulated with a metal shell. Mechanical performance was tested before and after encapsulation, which resulted in varying degrees of reduction in mechanical quality factor and resonant frequency. The acoustic performance of the ACUT was compared to commercial transducers, and the results indicated 2.72 and 1.88 dB higher transmitting and receiving sensitivities than those of commercial transducers. The received amplitude increased by an additional 22% by adding an electrical impedance matching circuit between the excitation source and the transmitting transducer. High-temperature experiments revealed a linear decrease in the received signal amplitude with increasing temperature. The ACUT is finally calibrated for ambient flow rate calibration and high-temperature flow velocity calibration, respectively. The ambient flow rate calibration exhibited an error of only 1.9%, while the high-temperature flow velocity calibration demonstrated that temperature has a minimal impact on the precision of the ACUT.

Authors

Institutions

Publication Details

Journal
IEEE Sensors Journal
Published
2024-08-12
DOI
https://doi.org/10.1109/jsen.2024.3434449
Citations
7
Primary Topic
Flow Measurement and Analysis
Type
article
Field-Weighted Citation Impact
1.88
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimized Design of High-Temperature Resistant Air-Coupled Ultrasonic Transducer (ACUT) Based on Single Acoustic Matching Layer

Liang Zhang, Xinhu Xu, Hulin Guo, Lingcai Kong et al.
7 citations
IEEE Sensors Journal
Flow Measurement and Analysis
1.88
article

Optimized Design of High-Temperature Resistant Air-Coupled Ultrasonic Transducer (ACUT) Based on Single Acoustic Matching Layer

Liang Zhang, Xinhu Xu, Hulin Guo, Lingcai Kong, Lei Li, Jinliang Zang, Wanglong Shi, Xiaojie Wang
article en
7 citations

Abstract

A high-temperature resistant air-coupled ultrasonic transducer (ACUT) was optimally designed with a single acoustic matching layer structure. The matching layer was prepared with an organosilicon substrate instead of traditional epoxy resin mixed with hollow glass microspheres. Acoustic performance tests showed that the matching layer achieved an acoustic impedance of 1.06 MRayl and an acoustic attenuation coefficient of 0.47 dB/mm. The ACUT was clamped with a prestressed bolt structure instead of an adhesive layer and encapsulated with a metal shell. Mechanical performance was tested before and after encapsulation, which resulted in varying degrees of reduction in mechanical quality factor and resonant frequency. The acoustic performance of the ACUT was compared to commercial transducers, and the results indicated 2.72 and 1.88 dB higher transmitting and receiving sensitivities than those of commercial transducers. The received amplitude increased by an additional 22% by adding an electrical impedance matching circuit between the excitation source and the transmitting transducer. High-temperature experiments revealed a linear decrease in the received signal amplitude with increasing temperature. The ACUT is finally calibrated for ambient flow rate calibration and high-temperature flow velocity calibration, respectively. The ambient flow rate calibration exhibited an error of only 1.9%, while the high-temperature flow velocity calibration demonstrated that temperature has a minimal impact on the precision of the ACUT.

IEEE Sensors JournalVol. 24(17)
National Institute of Metrology (CN), Hebei University (CN)
Affordable and clean energy
Openalex Percentile: Top 16%
Flow Measurement and Analysis
1.88
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.