Feasibility of capturing vessel expansion with 4D‐CTA: Phantom study to determine reproducibility, spatial and temporal resolution
In a phantom setting, 4D CTA can be used to reliably capture reproducible tube diameter changes exceeding 0.30 mm. Low pulsation frequencies (40 or 50 bpm) provide an accurate measurement of the maximum tube diameter change.