Removal of leads broken during extraction: A comparison of different approaches and tools
(1) Effectiveness of fractured lead removal is satisfactory: entire BLFs were removed in 61.4% (total procedural success-57.5%, was lower because five major complications occurred) and BLF length was significantly reduced in 26.0%. (2) Among the broken leads, leads with a long stay in the patient (16.3 years on average), passive leads (97.6%) and pacemaker leads 92.1% are significantly more common, but not ICD leads (only 7.9% of lead fractures) compared to TLE without lead fractures. (3) Broken lead removal (superior approach) using a CS access sheath as a "subclavian workstation" for continuation of dilatation with conventional tools deserves attention. (4) Lead fracture management should become an integral part of training in transvenous lead extraction.
Authors
- Dorota Nowosielecka (ORCID: https://orcid.org/0000-0002-2441-3876)
- Sebastian Sawonik (ORCID: https://orcid.org/0000-0003-2738-5214)
- Jarosław Kosior (ORCID: https://orcid.org/0000-0002-8526-0500)
- Wojciech Jacheć (ORCID: https://orcid.org/0000-0002-1091-9788)
- Andrzej Kutarski (ORCID: https://orcid.org/0000-0003-3990-1028)
- Łukasz Tułecki
- Paweł Stefańczyk (ORCID: https://orcid.org/0000-0002-4106-3804)
- Radosław Pietura (ORCID: https://orcid.org/0000-0001-9233-0637)
- Marek Czakowski
Institutions
- Medical University of Lublin (PL)
- Medical University of Silesia (PL)
- Radom College (PL)
- John Paul II Hospital (PL)
Publication Details
- Journal
- Journal of Cardiovascular Electrophysiology
- Published
- 2024-08-11
- DOI
- https://doi.org/10.1111/jce.16398
- Citations
- 1
- Primary Topic
- Cardiac pacing and defibrillation studies
- Type
- article
- Field-Weighted Citation Impact
- 0.40