A novel etiology for automatic mode switching and ventricular noise reversion alerts: pulsed field ablation

Artifactual recordings on cardiac implantable electronic devices (CIEDs) may be physiologic oversensing or a harbinger of lead failure with implications of requiring lead revision. 1,2oise reversion is an algorithm that causes the pacing functionality of devices to switch to asynchronous pacing when there is repetitive sensing at high rates. 3With the recent approval from the U.S. Food and Drug Administration for two pulsed-field ablation (PFA) systems and concurrent increased uptake of the technology for catheter ablation of atrial fibrillation, artifactual recordings specific to PFA might be recorded from a CIED, warranting distinction from other environmental sources of oversensing and noise reversion alerts.A 76-year-old male patient with a medical history of ischemic cardiomyopathy and prior substrate modification for a VT originating from left ventricular anterior/anteroseptal scar developed persistent atrial fibrillation.He had a dual-chamber implantable cardioverter defibrillator (Gallant DR, Abbott, Chicago, IL) programmed DDDR (70-100) with automatic mode switching to DDIR programmed at an atrial rate of 160 bpm.He presented for catheter ablation using the pentaspline PFA catheter (Farawave, Boston Scientific, Marlborough, MA).Device therapies were turned off before the procedure and a preprocedure device interrogation showed normal device function and no significant events other than persistent atrial fibrillation.Upon interrogation of the device postprocedure, two automatic mode switching episodes (Figure 1) and four noise reversion alerts (Figure 2) were noted on routine post-ablation device interrogation.During delivery of each lesion, unsynchronized therapeutic waveforms in a hierarchical set of microsecond scale biphasic pulses are delivered. 4With the increased use of PFA nationally and internationally, patients with CIEDs are likely to record such artifactual tracings.It is valuable for all clinicians interacting with CIEDs to appreciate these patterns of oversensing when investigating device alerts during interrogations, particularly as these findings are likely to become more prevalent with continued uptake of PFA technologies.

Authors

Institutions

Publication Details

Journal
HeartRhythm Case Reports
Published
2024-08-10
DOI
https://doi.org/10.1016/j.hrcr.2024.08.006
Citations
4
Primary Topic
Cardiac electrophysiology and arrhythmias
Type
article
Field-Weighted Citation Impact
1.61
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A novel etiology for automatic mode switching and ventricular noise reversion alerts: pulsed field ablation

Marc A. Miller, Jacob S. Koruth, Vivek Y. Reddy, Joshua Lampert
4 citations
HeartRhythm Case Reports
Cardiac electrophysiology and arrhythmias
1.61
article

A novel etiology for automatic mode switching and ventricular noise reversion alerts: pulsed field ablation

Marc A. Miller, Jacob S. Koruth, Vivek Y. Reddy, Joshua Lampert
article en
4 citations

Abstract

Artifactual recordings on cardiac implantable electronic devices (CIEDs) may be physiologic oversensing or a harbinger of lead failure with implications of requiring lead revision. 1,2oise reversion is an algorithm that causes the pacing functionality of devices to switch to asynchronous pacing when there is repetitive sensing at high rates. 3With the recent approval from the U.S. Food and Drug Administration for two pulsed-field ablation (PFA) systems and concurrent increased uptake of the technology for catheter ablation of atrial fibrillation, artifactual recordings specific to PFA might be recorded from a CIED, warranting distinction from other environmental sources of oversensing and noise reversion alerts.A 76-year-old male patient with a medical history of ischemic cardiomyopathy and prior substrate modification for a VT originating from left ventricular anterior/anteroseptal scar developed persistent atrial fibrillation.He had a dual-chamber implantable cardioverter defibrillator (Gallant DR, Abbott, Chicago, IL) programmed DDDR (70-100) with automatic mode switching to DDIR programmed at an atrial rate of 160 bpm.He presented for catheter ablation using the pentaspline PFA catheter (Farawave, Boston Scientific, Marlborough, MA).Device therapies were turned off before the procedure and a preprocedure device interrogation showed normal device function and no significant events other than persistent atrial fibrillation.Upon interrogation of the device postprocedure, two automatic mode switching episodes (Figure 1) and four noise reversion alerts (Figure 2) were noted on routine post-ablation device interrogation.During delivery of each lesion, unsynchronized therapeutic waveforms in a hierarchical set of microsecond scale biphasic pulses are delivered. 4With the increased use of PFA nationally and internationally, patients with CIEDs are likely to record such artifactual tracings.It is valuable for all clinicians interacting with CIEDs to appreciate these patterns of oversensing when investigating device alerts during interrogations, particularly as these findings are likely to become more prevalent with continued uptake of PFA technologies.

HeartRhythm Case ReportsVol. 10(11)
Icahn School of Medicine at Mount Sinai (US)
Openalex Percentile: Top 15%
Cardiac electrophysiology and arrhythmias
1.61
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.