Effects of Patient Dose Reduction Efforts on Image Quality for Thoracic CT in A Moroccan Hospital

Thoracic CT imaging is a strongly recommended means of medical imaging. It is accompanied by a high exposure to radiation due to the number of examinations carried out, which requires efforts to reduce the CT dose index (CTDI) while trying to preserve image quality. To this end, this study proposed the possibility of introducing two new imaging protocols for chest tomography. A 16-slice HITACHI SUPIA CT scanner and two phantoms were used to investigate CTDIvol and image quality; the first phantom was made of PMMA and the other of AAPM model 610. Three tube voltages were studied by varying the intensity of the tube current (mA): 120 kVp (120-160-210-230) mA, 100 kVp (160-200-240-290) mA, and 80 kVp (230-260-300-350) mA. The values for noise uniformity and accuracy, contrast-to-noise ratio (CNR), and spatial resolution (SR) were determined using IndoQCT c22a. 92 software. The first thoracic protocol proposed with 100 kVp compared to 120 kVp resulted in a 27.51 % reduction in CTDIvol, a 20 % increase in mA, and a 19.50 % increase in noise. The CNR showed a slight regression of 23.08 %. For the second scan procedure at 80 kVp, the CTDIvol was reduced by 53.32 %, while noise was increased by 53.95 %. There was no statistically significant difference in CNR and SR (p > 0.05) when kVp and mA were reduced compared to the routine protocol. It is suggested that it is possible to adopt two new acquisition protocols at 100 kVp and 80 kVp while reducing the patient exposure dose (CTDIvol) by 28 % and 54 % and taking into account the effect of varying these parameters on image quality. Their choices must be made by integrating and considering clinical issues and a good understanding of the pathophysiology and imaging results of the suspected condition. Consequently, radiologists and technicians should always take a part in improving imaging practices in such a way as to make more effective use of radiation.

Authors

Institutions

Publication Details

Journal
Atom Indonesia
Published
2024-08-08
DOI
https://doi.org/10.55981/aij.2024.1411
Citations
2
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.81
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effects of Patient Dose Reduction Efforts on Image Quality for Thoracic CT in A Moroccan Hospital

Mohammed Talbi, Zouhir Saga, M’hamed El Mansouri, A. Rezzouk et al.
2 citations
Atom Indonesia
Radiation Dose and Imaging
0.81
article

Effects of Patient Dose Reduction Efforts on Image Quality for Thoracic CT in A Moroccan Hospital

Mohammed Talbi, Zouhir Saga, M’hamed El Mansouri, A. Rezzouk, L. Belaroussi, Alain Rahmouni
article en
2 citations

Abstract

Thoracic CT imaging is a strongly recommended means of medical imaging. It is accompanied by a high exposure to radiation due to the number of examinations carried out, which requires efforts to reduce the CT dose index (CTDI) while trying to preserve image quality. To this end, this study proposed the possibility of introducing two new imaging protocols for chest tomography. A 16-slice HITACHI SUPIA CT scanner and two phantoms were used to investigate CTDIvol and image quality; the first phantom was made of PMMA and the other of AAPM model 610. Three tube voltages were studied by varying the intensity of the tube current (mA): 120 kVp (120-160-210-230) mA, 100 kVp (160-200-240-290) mA, and 80 kVp (230-260-300-350) mA. The values for noise uniformity and accuracy, contrast-to-noise ratio (CNR), and spatial resolution (SR) were determined using IndoQCT c22a. 92 software. The first thoracic protocol proposed with 100 kVp compared to 120 kVp resulted in a 27.51 % reduction in CTDIvol, a 20 % increase in mA, and a 19.50 % increase in noise. The CNR showed a slight regression of 23.08 %. For the second scan procedure at 80 kVp, the CTDIvol was reduced by 53.32 %, while noise was increased by 53.95 %. There was no statistically significant difference in CNR and SR (p > 0.05) when kVp and mA were reduced compared to the routine protocol. It is suggested that it is possible to adopt two new acquisition protocols at 100 kVp and 80 kVp while reducing the patient exposure dose (CTDIvol) by 28 % and 54 % and taking into account the effect of varying these parameters on image quality. Their choices must be made by integrating and considering clinical issues and a good understanding of the pathophysiology and imaging results of the suspected condition. Consequently, radiologists and technicians should always take a part in improving imaging practices in such a way as to make more effective use of radiation.

Atom IndonesiaVol. 50(2)
Université Ibn-Tofail (MA), Sidi Mohamed Ben Abdellah University (MA)
Openalex Percentile: Top 26%
Radiation Dose and Imaging
0.81
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.