Unexpected discrepancy between the planned gating window and the actual irradiated gating window in phase-based gating radiotherapy during regular respiration

We observed unexpected gating window discrepancies in asymmetric respiratory patterns. Therefore, the present study was conducted to test the hypothesis that asymmetric breathing causes this discrepancy. Six regular respiratory patterns were generated using the QUASAR software. 4-dimensional computed tomography (4DCT) was acquired with each breathing pattern using a QUASAR respiratory motion phantom with a 1.2 cm spherical acrylic ball embedded in a cedar insert. A Real-time Position Management system was used for respiratory signal acquisition during the 4DCT. Each treatment plan was created using gating windows ranging from 30% to 70%. TrueBeam STx was used for gated beam delivery, and the stored waveform data files and gated kV images were analyzed after irradiation. Two different concepts of phase were used to analyze the waveform data files. The respiratory phase defined prospectively by the TrueBeam gating system during treatment was called “phase pro ”, and the phase defined retrospectively by waveform data analysis after treatment was called “phase retro ". The phase pro values were close to the planned gating window (30% or 70%) for in all breathing patterns. However, the phase retro values differed from the planned gating window in the asymmetric patterns. The gating window was either pulled forward or pushed back from the treatment plan depending on the direction of the asymmetry. Furthermore, this discrepancy increased as the asymmetry of the respiratory pattern increased. In some asymmetric breathing models, the tumor location on the gated kV image was outside the internal target volume contour. This study confirmed that the gating window discrepancy between 4DCT and the TrueBeam gating system occurs with asymmetric respiration pattern, even with regular breathing. Therefore, if a patient's breathing is not exactly symmetrical, appropriate measures should be taken against this unexpected systematic error.

Authors

Institutions

Publication Details

Journal
Journal of Radiation Research and Applied Sciences
Published
2024-08-09
DOI
https://doi.org/10.1016/j.jrras.2024.101072
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Unexpected discrepancy between the planned gating window and the actual irradiated gating window in phase-based gating radiotherapy during regular respiration

Hyoung Uk Je, Dae Hyun Kim, Young Eun Ko
Journal of Radiation Research and Applied Sciences
Advanced Radiotherapy Techniques
article

Unexpected discrepancy between the planned gating window and the actual irradiated gating window in phase-based gating radiotherapy during regular respiration

Hyoung Uk Je, Dae Hyun Kim, Young Eun Ko
article en

Abstract

We observed unexpected gating window discrepancies in asymmetric respiratory patterns. Therefore, the present study was conducted to test the hypothesis that asymmetric breathing causes this discrepancy. Six regular respiratory patterns were generated using the QUASAR software. 4-dimensional computed tomography (4DCT) was acquired with each breathing pattern using a QUASAR respiratory motion phantom with a 1.2 cm spherical acrylic ball embedded in a cedar insert. A Real-time Position Management system was used for respiratory signal acquisition during the 4DCT. Each treatment plan was created using gating windows ranging from 30% to 70%. TrueBeam STx was used for gated beam delivery, and the stored waveform data files and gated kV images were analyzed after irradiation. Two different concepts of phase were used to analyze the waveform data files. The respiratory phase defined prospectively by the TrueBeam gating system during treatment was called “phase pro ”, and the phase defined retrospectively by waveform data analysis after treatment was called “phase retro ". The phase pro values were close to the planned gating window (30% or 70%) for in all breathing patterns. However, the phase retro values differed from the planned gating window in the asymmetric patterns. The gating window was either pulled forward or pushed back from the treatment plan depending on the direction of the asymmetry. Furthermore, this discrepancy increased as the asymmetry of the respiratory pattern increased. In some asymmetric breathing models, the tumor location on the gated kV image was outside the internal target volume contour. This study confirmed that the gating window discrepancy between 4DCT and the TrueBeam gating system occurs with asymmetric respiration pattern, even with regular breathing. Therefore, if a patient's breathing is not exactly symmetrical, appropriate measures should be taken against this unexpected systematic error.

Journal of Radiation Research and Applied SciencesVol. 17(4)
Ulsan University Hospital (KR)
Zero hunger
Openalex Percentile: Top 87%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.