Radiation Dose during Digital Subtraction Angiography of the Brain—The Influence of Examination Parameters and Patient Factors on the Dose

Cerebral vascular angiography, or digital subtraction angiography (DSA), is essential for diagnosing neurological conditions but poses radiation risks. This study aims to analyze the impact of examination parameters and patient characteristics on the radiation dose received during DSA to optimize safety and minimize exposure. A retrospective analysis of 251 DSA procedures using the GE Innova IGS 630 dual-plane instrument was conducted. Data on dose area product (DAP) and air kerma (KERMA), along with patient and examination details, were collected. Statistical analyses, including Mann-Whitney, Kruskal-Wallis, and Spearman rank correlation tests, assessed the relationships between variables and radiation dose outcomes. Significant correlations were found between the sides examined (left, right, or both) and DAP (<i>p</i> < 0.0001) and KERMA (<i>p</i> < 0.0001) values, with bilateral studies showing the highest values. The post hoc Dunn tests showed that the 'L + P' group significantly differs from both the right group (<i>p</i> < 0.0001 and the left group (<i>p</i> < 0.0001). There is no significant difference between the 'P' group and the 'L' group (<i>p</i>-value = 0.53). These results suggest that the right and left (both) group have unique KERMA mGy values compared to the other two groups. A strong correlation (rS = 0.87) existed between DAP and KERMA. The number of projections significantly impacted radiation dose (rS = 0.61). Tube parameters (kV and mA) and skull size had low correlations with DAP and KERMA. Optimizing imaging protocols and individualizing parameters can significantly enhance patient safety and diagnostic efficacy while also reducing occupational exposure for medical staff.

Authors

Institutions

Publication Details

Journal
Brain Sciences
Published
2024-08-09
DOI
https://doi.org/10.3390/brainsci14080799
Citations
1
Primary Topic
Cerebrovascular and Carotid Artery Diseases
Type
article
Field-Weighted Citation Impact
0.42

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Radiation Dose during Digital Subtraction Angiography of the Brain—The Influence of Examination Parameters and Patient Factors on the Dose

Sandra Modlińska, Marcin Rojek, Miłosz Zbroszczyk, Piotr Dudek et al.
1 citations
Brain Sciences
Cerebrovascular and Carotid Artery Diseases
0.42
article

Radiation Dose during Digital Subtraction Angiography of the Brain—The Influence of Examination Parameters and Patient Factors on the Dose

Sandra Modlińska, Marcin Rojek, Miłosz Zbroszczyk, Piotr Dudek, Łukasz Czogalik, Jakub Kufel, Michał Janik
article en
1 citations

Abstract

Cerebral vascular angiography, or digital subtraction angiography (DSA), is essential for diagnosing neurological conditions but poses radiation risks. This study aims to analyze the impact of examination parameters and patient characteristics on the radiation dose received during DSA to optimize safety and minimize exposure. A retrospective analysis of 251 DSA procedures using the GE Innova IGS 630 dual-plane instrument was conducted. Data on dose area product (DAP) and air kerma (KERMA), along with patient and examination details, were collected. Statistical analyses, including Mann-Whitney, Kruskal-Wallis, and Spearman rank correlation tests, assessed the relationships between variables and radiation dose outcomes. Significant correlations were found between the sides examined (left, right, or both) and DAP (<i>p</i> < 0.0001) and KERMA (<i>p</i> < 0.0001) values, with bilateral studies showing the highest values. The post hoc Dunn tests showed that the 'L + P' group significantly differs from both the right group (<i>p</i> < 0.0001 and the left group (<i>p</i> < 0.0001). There is no significant difference between the 'P' group and the 'L' group (<i>p</i>-value = 0.53). These results suggest that the right and left (both) group have unique KERMA mGy values compared to the other two groups. A strong correlation (rS = 0.87) existed between DAP and KERMA. The number of projections significantly impacted radiation dose (rS = 0.61). Tube parameters (kV and mA) and skull size had low correlations with DAP and KERMA. Optimizing imaging protocols and individualizing parameters can significantly enhance patient safety and diagnostic efficacy while also reducing occupational exposure for medical staff.

Brain SciencesVol. 14(8)
Medical University of Silesia (PL)
Śląski Uniwersytet Medyczny
Openalex Percentile: Top 34%
Cerebrovascular and Carotid Artery Diseases
0.42
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.