Calibration of Aseptic Loosening Simulation for Coatings Osteoinductive Effect

The risk of aseptic loosening in cementless hip stems can be reduced by improving osseointegration with osteoinductive coatings favoring long-term implant stability. Osseointegration is usually evaluated in vivo studies, which, however, do not reproduce the mechanically driven adaptation process. This study aims to develop an in silico model to predict implant osseointegration and the effect of induced micromotion on long-term stability, including a calibration of the material osteoinductivity with conventional in vivo studies. A Finite Element model of the tibia implanted with pins was generated, exploiting bone-to-implant contact measures of cylindrical titanium alloys implanted in rabbits' tibiae. The evolution of the contact status between bone and implant was modeled using a finite state machine, which updated the contact state at each iteration based on relative micromotion, shear and tensile stresses, and bone-to-implant distance. The model was calibrated with in vivo data by identifying the maximum bridgeable gap. Afterward, a push-out test was simulated to predict the axial load that caused the macroscopic mobilization of the pin. The bone-implant bridgeable gap ranged between 50 μm and 80 μm. Predicted push-out strength ranged from 19 N to 21 N (5.4 MPa-3.4 MPa) depending on final bone-to-implant contact. Push-out strength agrees with experimental measurements from a previous animal study (4 ± 1 MPa), carried out using the same implant material, coated, or uncoated. This method can partially replace in vivo studies and predict the long-term stability of cementless hip stems.

Authors

Institutions

Publication Details

Journal
Annals of Biomedical Engineering
Published
2024-08-09
DOI
https://doi.org/10.1007/s10439-024-03588-9
Citations
2
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
1.15

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Calibration of Aseptic Loosening Simulation for Coatings Osteoinductive Effect

Melania Maglio, Antonino A. La Mattina, Sara Oliviero, Milena Fini et al.
2 citations
Annals of Biomedical Engineering
Orthopaedic implants and arthroplasty
1.15
article

Calibration of Aseptic Loosening Simulation for Coatings Osteoinductive Effect

Melania Maglio, Antonino A. La Mattina, Sara Oliviero, Milena Fini, Lúcia Martini, Marco Viceconti, Sofia Baroni
article en
2 citations

Abstract

The risk of aseptic loosening in cementless hip stems can be reduced by improving osseointegration with osteoinductive coatings favoring long-term implant stability. Osseointegration is usually evaluated in vivo studies, which, however, do not reproduce the mechanically driven adaptation process. This study aims to develop an in silico model to predict implant osseointegration and the effect of induced micromotion on long-term stability, including a calibration of the material osteoinductivity with conventional in vivo studies. A Finite Element model of the tibia implanted with pins was generated, exploiting bone-to-implant contact measures of cylindrical titanium alloys implanted in rabbits' tibiae. The evolution of the contact status between bone and implant was modeled using a finite state machine, which updated the contact state at each iteration based on relative micromotion, shear and tensile stresses, and bone-to-implant distance. The model was calibrated with in vivo data by identifying the maximum bridgeable gap. Afterward, a push-out test was simulated to predict the axial load that caused the macroscopic mobilization of the pin. The bone-implant bridgeable gap ranged between 50 μm and 80 μm. Predicted push-out strength ranged from 19 N to 21 N (5.4 MPa-3.4 MPa) depending on final bone-to-implant contact. Push-out strength agrees with experimental measurements from a previous animal study (4 ± 1 MPa), carried out using the same implant material, coated, or uncoated. This method can partially replace in vivo studies and predict the long-term stability of cementless hip stems.

Annals of Biomedical EngineeringVol. 53(1)
Istituto Ortopedico Rizzoli (IT), Istituti di Ricovero e Cura a Carattere Scientifico (IT), University of Bologna (IT)
European Commission, Università di Bologna, Istituto Ortopedico Rizzoli di Bologna, Regione Emilia-Romagna, Horizon 2020 Framework Programme
Openalex Percentile: Top 21%
Orthopaedic implants and arthroplasty
1.15
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.