Effect of single-photon emission computed tomography acquisition method and sampling angles on image quality and quantitative accuracy in xSPECT-reconstructed images
In xSPECT-reconstructed images, the difference in sampling angle has a greater impact on image quality and quantitativity than the difference in the acquisition method. For tests in which uniformity is important, a larger sampling angle is recommended.