Screening Mammography Diagnostic Reference Level System According to Compressed Breast Thickness: Dubai Health

Screening mammography is considered to be the most effective means for the early detection of breast cancer. However, epidemiological studies suggest that longitudinal exposure to screening mammography may raise breast cancer radiation-induced risk, which begs the need for optimization and internal auditing. The present work aims to establish a comprehensive well-structured Diagnostic Reference Level (DRL) system that can be confidently used to highlight healthcare centers in need of urgent action, as well as cases exceeding the dose notification level. Screening mammographies from a total of 2048 women who underwent screening mammography at seven different healthcare centers were collected and retrospectively analyzed. The typical DRL for each healthcare center was established and defined as per (A) bilateral image view (left craniocaudal (LCC), right craniocaudal (RCC), left mediolateral oblique (LMLO), and right mediolateral oblique (RMLO)) and (B) structured compressed breast thickness (CBT) criteria. Following this, the local DRL value was established per the bilateral image views for each CBT group. Screening mammography data from a total of 8877 images were used to build this comprehensive DRL system (LCC: 2163, RCC: 2206, LMLO: 2288, and RMLO: 2220). CBTs were classified into eight groups of <20 mm, 20-29 mm, 30-39 mm, 40-49 mm, 50-59 mm, 60-69 mm, 70-79 mm, 80-89 mm, and 90-110 mm. Using the Kruskal-Wallis test, significant dose differences were observed between all seven healthcare centers offering screening mammography. The local DRL values defined per bilateral image views for the CBT group 60-69 mm were (1.24 LCC, 1.23 RCC, 1.34 LMLO, and 1.32 RMLO) mGy. The local DRL defined per bilateral image view for a specific CBT highlighted at least one healthcare center in need of optimization. Such comprehensive DRL system is efficient, easy to use, and very clinically effective.

Authors

Institutions

Publication Details

Journal
Journal of Imaging
Published
2024-08-05
DOI
https://doi.org/10.3390/jimaging10080188
Citations
6
Primary Topic
Digital Radiography and Breast Imaging
Type
article
Field-Weighted Citation Impact
2.51
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Screening Mammography Diagnostic Reference Level System According to Compressed Breast Thickness: Dubai Health

Entesar Z. Dalah, Maryam K. Alkaabi, Nisha A. Antony, Hashim M. Al-Awadhi
6 citations
Journal of Imaging
Digital Radiography and Breast Imaging
2.51
article

Screening Mammography Diagnostic Reference Level System According to Compressed Breast Thickness: Dubai Health

Entesar Z. Dalah, Maryam K. Alkaabi, Nisha A. Antony, Hashim M. Al-Awadhi
article en
6 citations

Abstract

Screening mammography is considered to be the most effective means for the early detection of breast cancer. However, epidemiological studies suggest that longitudinal exposure to screening mammography may raise breast cancer radiation-induced risk, which begs the need for optimization and internal auditing. The present work aims to establish a comprehensive well-structured Diagnostic Reference Level (DRL) system that can be confidently used to highlight healthcare centers in need of urgent action, as well as cases exceeding the dose notification level. Screening mammographies from a total of 2048 women who underwent screening mammography at seven different healthcare centers were collected and retrospectively analyzed. The typical DRL for each healthcare center was established and defined as per (A) bilateral image view (left craniocaudal (LCC), right craniocaudal (RCC), left mediolateral oblique (LMLO), and right mediolateral oblique (RMLO)) and (B) structured compressed breast thickness (CBT) criteria. Following this, the local DRL value was established per the bilateral image views for each CBT group. Screening mammography data from a total of 8877 images were used to build this comprehensive DRL system (LCC: 2163, RCC: 2206, LMLO: 2288, and RMLO: 2220). CBTs were classified into eight groups of <20 mm, 20-29 mm, 30-39 mm, 40-49 mm, 50-59 mm, 60-69 mm, 70-79 mm, 80-89 mm, and 90-110 mm. Using the Kruskal-Wallis test, significant dose differences were observed between all seven healthcare centers offering screening mammography. The local DRL values defined per bilateral image views for the CBT group 60-69 mm were (1.24 LCC, 1.23 RCC, 1.34 LMLO, and 1.32 RMLO) mGy. The local DRL defined per bilateral image view for a specific CBT highlighted at least one healthcare center in need of optimization. Such comprehensive DRL system is efficient, easy to use, and very clinically effective.

Journal of ImagingVol. 10(8)
Dubai Health Authority (AE), Dubai Hospital (AE), Mohammed Bin Rashid University of Medicine and Health Sciences (AE)
Good health and well-being
Openalex Percentile: Top 10%
Digital Radiography and Breast Imaging
2.51
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.