Research on calibration of the blood transfusion and infusion heating device
This article studies the calibration method of blood transfusion and infusion heating device, which will provide effective reference for the metrological performance evaluation of the device. Calibration methods for key technical parameters of the device, such as temperature indication error, temperature fluctuation, and temperature uniformity, are designed. 7 different types of blood transfusion and infusion heating devices widely used in the hospital are selected to evaluate the feasibility of the calibration method. The calibration results prove that the designed calibration method for blood transfusion and infusion heating device can improve the metrological traceability system of the device, ensure the safety and effectiveness of its performance, and reduce clinical risks.
Authors
- Han Luo (ORCID: https://orcid.org/0000-0002-2009-9365)
- Ting-Ting Ren (ORCID: https://orcid.org/0000-0002-1108-7450)
- Biao Gao (ORCID: https://orcid.org/0000-0001-9241-9447)
- Huijuan Wang (ORCID: https://orcid.org/0000-0003-3271-9709)
- Jing Huang (ORCID: https://orcid.org/0009-0007-5194-9960)
Institutions
- Army Medical University (CN)
- Chongqing Metrology Quality Inspection and Research Institute (CN)
Publication Details
- Published
- 2024-08-05
- DOI
- https://doi.org/10.1117/12.3038275
- Primary Topic
- Intravenous Infusion Technology and Safety
- Type
- article
- Field-Weighted Citation Impact
- 0.00